EL3020 belongs to the EasyLine EL3000 series. It is the 19-inch rack-mount version in ABB’s environmental analysis product line (the wall-mounted version of the same series is EL3040).
基本仕様
取り付け: 19-inch rack, 3Uの高さ
寸法: 483 mm (W) × 357 mm (D) × 132 mm (H)
保護クラス: IP20; IP40 for CEMS version
電源: 100~240 VAC
構造: A single cabinet can be fitted with a maximum of 2 analysis modules, measuring up to 5 gas components simultaneously
オプションのアクセサリ: Built-in sample gas pump, 電磁弁, フィルター, flow sensor
起源: ドイツ
Available Measuring Modules (mix and match supported)
| モジュールモデル | 原理 | Measurable Gas |
| Uras26 | NDIR Non-Dispersive Infrared | CO, CO₂, CH₄, SO₂, いいえ |
| Limas23 | UV Ultraviolet Absorption | いいえ, NO₂, SO₂ |
| Magnos28 | Paramagnetic Oxygen | O₂ |
| Magnos27 | Thermomagnetic Oxygen | O₂ |
| ZO23 | Zirconia Trace Oxygen | Trace O₂ |
| Caldos27 | Thermal Conductivity | H₂, Ar and other binary gas mixtures |
| Fidas24 | FID Flame Ionization Detection | Total Hydrocarbons / HC |
インターフェース & コミュニケーション
イーサネット (内蔵Webサーバー)
プロフィバスDP
Modbus
4…20 mA analog output
Digital DI/DO (最大. 12 から + 12 する)
Local LCD with 5-button operation, Chinese display supported; ECT configuration software for advanced parameter setup
主な特長
- Built-in calibration cell: Clean air can be used for zero calibration in most working conditions, eliminating the need for permanent standard gas cylinders; automatic calibration supported
- Built-in self-diagnostics, optional QAL3 (EN14181) certification for flue gas continuous monitoring, suitable for CEMS emission monitoring
- ABB Ability digital function: Quick QR code access for diagnostics, remote support, status alerts, cluster monitoring on cloud platform
- モジュラー設計: Modules can be added or replaced later without replacing the whole unit
代表的な用途
Flue gas emission monitoring (CEMS) for power plants, waste incineration, metallurgy and boilers
Biogas, landfill gas and fermentation gas analysis
Blast furnace gas and reformer process gas analysis
Gas production, air separation, inert gas purity and trace oxygen monitoring
Combustion monitoring for engines and gas turbines
スペアパーツ & 税関情報
HSコード: 90279000 (same classification for gas analyzer spare parts and complete unit)
Common consumables: Sample filter element, pump diaphragm, FID flame nozzle, gas capillary, FFKM sealing O-ring, light source
Quick Comparison with EL3040
EL3020: Rack-mounted, IP20, installed inside control cabinet
EL3040: 壁掛け, IP65 purge type, field local installation
Operating Instructions for ABB EL3020 EasyLine Extractive Analyzer
The front panel adopts 5-button operation: わかりました (確認する / enter menu), 上 / 下 / 左 / right navigation keys. The screen supports Chinese language. The operation modes include measurement view, menu setting, 校正と診断. It can also be operated remotely via ECT configuration software on PC or built-in web page.
Important precondition: EL3020 is 19-inch rack-mounted with IP20, no cabinet purge. Remove the yellow protective cap of the pressure sensor before power-on. Check gas circuit, heat tracing, フィルター, pump and flow in advance; no condensation or leakage allowed.
- Power-on Operation (First Startup / Restart after Long Shutdown)
- Pre-startup inspection
Gas circuit: Sample gas, zero gas and span gas pipelines connected properly, switching valves functional; filter cartridge intact; flow rate set to rated value (generally 30–60 L/h)
Electrical circuit: 100–240 VAC power supply, 4–20mA, DI/DO and communication wiring correct
Pressure sensor: Remove yellow plastic cap and connect T-type pressure compensation pipeline as required
- Pipeline pre-purge (mandatory for new commissioning or after maintenance)
Purge the whole sample gas loop with nitrogen or clean zero gas to remove residual impurities and corrosive gas and avoid contamination of optical modules. Purge time depends on pipeline length, normally ≥20 min.
- Power on and warm-up
Switch on power; the analyzer starts automatically. Warm-up duration varies by module
Uras26 infrared: 約. 30 分
Limas23 UV, Magnos paramagnetic oxygen: 30–60 min
Fidas24 FID: ≥60 min, requires hydrogen and combustion air
ZO23 zirconia trace oxygen: longer thermostatic stabilization time
- After warm-up completion, the main measurement interface appears, showing component concentration, ユニット, 流れ, status and alarm information. Sample gas can be fed in for continuous measurement.
ヒント: Versions with built-in calibration cell (Uras26) can use clean air for daily zero calibration without continuous zero gas cylinder supply.
- Panel Buttons & Menu Access
Normal measurement screen: Real-time concentration, status, flow and temperature
Press【OK】to enter main menu; select items with up/down keys, confirm with OK, modify parameters with left/right keys, exit with return key
Main menu hierarchy:
- Measurement View
- 較正 (most frequently used)
- 診断 / アラーム
- 構成: 範囲, ユニット, アラームしきい値, アナログ出力, コミュニケーション, 時間, language
- Log: alarm log, calibration record, QAL3 report (オプション)
- Info: module version, シリアルナンバー
- Daily Operation Monitoring (Normal Continuous Measurement)
- Feed stable sample gas after pre-treatment, filtration and moisture removal before entering the analyzer
- Screen observation:
Stable readings for each component, no flashing fault codes
Flow within set range; low flow triggers alarm to protect gas cell
ステータスインジケーター: 緑色 = 通常; yellow = warning; 赤 = 故障
- Remote access:
Enter analyzer IP address in browser to open built-in web server for real-time data, 曲線, alarms and calibration records. Use ECT EasyLine Configuration Tool for advanced parameter setting via RS232 service port or Ethernet. TCT software is used for in-depth diagnosis of optical modules.
- 信号出力: 4–20mA, Modbus / Profibus transmit concentration, status and fault signals to PLC/DCS synchronously.
- 較正 (鍵: ゼロ / スパン)
原理: Zero first, then span; start calibration only after readings stabilize with gas flowing. Follow QAL3 cycle requirements for CEMS environmental applications.
4.1 Manual Zero Calibration
- Menu path: OK → Calibration → Zero Cal
- Select target channel (CO/SO₂/NO/O₂ etc.)
- Switch gas circuit to zero gas (high-purity N₂ or clean zero air free of target components). Keep gas flowing for 3–5 min until reading stable.
- Select【Start Zero Calibration】
- Wait for automatic completion; screen shows Zero OK and reading returns near zero
- Switch back to sample gas
4.2 Manual Span Calibration
- Menu path: OK → Calibration → Span Cal
- Select target channel and input nominal concentration on gas certificate
- Switch to standard span gas, keep gas flowing for 5–10 min until reading stable
- Start span calibration; the instrument automatically corrects slope
- Completion prompt Span OK, check deviation between reading and certified span gas concentration
- Switch back to sample gas for measurement
4.3 Automatic Calibration (Models with built-in calibration cell, recommended for long-term operation)
- メニュー: Calibration → Auto Cal for setting
- Set calibration cycle, 順序, automatic valve switching and alarm logging
- After enabling auto-calibration, the analyzer drives internal solenoid valves at scheduled time and completes zero/span via built-in calibration cell, no external span gas cylinder required (standard for Uras26)
- FID, Limas, ZO23 and Caldos modules do NOT support built-in calibration cell; external standard gas is mandatory.
Calibration Differences by Module
Uras26 (NDIR): Supports auto zero via built-in cell; span can use internal cell or external standard gas
Limas23 (紫外線): Generally calibrated with external standard gas
Magnos28 paramagnetic oxygen: Single-point calibration with clean air
Fidas24 FID THC: Zero with zero air, span with methane standard gas; stable hydrogen pressure required
ZO23 trace oxygen: Calibrated with reference air; pay attention to sealing against leakage
- アラーム & フォルトリセット
- Alarm checking: Menu Diagnosis for active alarms and historical alarm log with text description
Common alarms: 低流量, temperature abnormality, light source failure, pressure fault, calibration failure, FID flameout, over-range
- Reset after troubleshooting:
General warning: auto-recover after root cause eliminated
Latched fault: select【Reset Fault】in menu; some severe faults require power cycle
- 重要: After FID flameout, check hydrogen and combustion air, purge before re-ignition; avoid repeated ignition directly.
- パラメータ設定
- Range modification: Set upper/lower limit for each component; recalibration is recommended after range change
- 4–20 mA analog output: Set mA mapping range, fault current (3.5 ミリアンペア / 22 ミリアンペア)
- ディ/ドゥ: Custom digital input (external calibration trigger, インターロック) and digital output (故障, calibration in progress, high concentration alarm)
- コミュニケーション: Modbus TCP/RTU, Profibus address and baud rate; Modbus port is disabled by factory default and needs enabling via ECT
- 言語, time and password management (advanced parameters require engineer password)
- Shutdown Operation
- Short shutdown (within several hours): Cut sample gas supply, keep purge gas and power on with constant temperature, no power-off to prevent condensate contamination of gas cell
- Long shutdown (>1 day)
Purge gas circuit with zero gas for 5–10 min
For FID version: Cut hydrogen first, extinguish flame, continue purging FID with combustion air
Close sample gas and standard gas cylinders
Finally cut off analyzer power supply
Seal pipelines for dust prevention
- Daily Operation Notes
- Sample gas must be pre-treated: dewatering, dust removal and tar removal; condensate is strictly forbidden to enter the analyzer
- Record each calibration: 時間, span gas concentration and result, archive for CEMS QAL3 compliance
- 定期点検: filter element, pump diaphragm, flow and pressure; check FID nozzle and hydrogen pressure reducing valve regularly
- Avoid long-time impact of over-range or high-concentration corrosive gas which may damage light source and gas cell
- Troubleshooting for failed calibration: gas leakage, unstable flow, expired standard gas, water vapor interference or light source aging
- Brief ECT Software Operation on PC
- 繋がり: RS232 service port or Ethernet
- Read device parameters and back up configuration
- Modify advanced calibration parameters, curves and compensation coefficients (プレッシャー & 温度補償)
- Firmware upgrade and deep diagnosis (TCT software for optical module light source and gas cell diagnosis)
ABB EasyLine EL3000 Series Complete Model & 注文情報
重要な注意事項:
EL3000 does not have fixed single part numbers. It is a configurable system consisting of chassis + measuring module + I/O + gas accessories. The complete system code is a long string `24042-xxxxxxxxxx` generated according to configuration and cannot be fixed statically. Below are base chassis models, measuring module models, descriptions, sample system codes and spare part ordering numbers.
There are only two base chassis models in this series: EL3020 (19” rack, IP20) and EL3040 (wall-mounted IP65 purge type). EL3010 / EL3030 / EL3050 do not exist. EL3060 is an independent Magnos206 unit and NOT part of the EL3020 / EL3040 chassis family.
- Base Chassis Models (2 Types)
| Chassis Model | 取り付け | 保護 | 説明 |
| EL3020 | 19-inch rack, 3U | IP20 (IP40 for CEMS version) | Installed in control cabinet; optional built-in sample pump, solenoid valves and filter; 最大. 2 モジュール, まで 5 components |
| EL3040 | 壁掛け | IP65 (purge capable) | Field local installation; purge housing for dust and water protection; also supports max. 2 モジュール |
- All Compatible Measuring Modules for EL3000 Series
| モジュールモデル | 原理 | Measurable Gas | 特徴 |
| Uras26 | NDIR Non-Dispersive Infrared | CO, CO₂, SO₂, いいえ, CH₄, N₂O | With built-in calibration cell, mainstream for CEMS, まで 4 infrared components |
| Limas23 | UV Ultraviolet Absorption | SO₂, いいえ, NO₂ | Flue gas UV module; no built-in calibration cell, external standard gas required |
| Magnos206 | Paramagnetic Oxygen | O₂ | New paramagnetic oxygen, replacement for legacy Magnos28; range down to 0–2%Vol, supports 98–100% high oxygen |
| Magnos27 | Thermomagnetic Oxygen | O₂ | Thermomagnetic principle, for conventional oxygen, starting from 0–5%Vol |
| ZO23 | Zirconia Trace Oxygen | Trace O₂ | ppb-level trace oxygen, for inert gas and high-purity gas monitoring |
| Caldos27 | Thermal Conductivity | H₂, Ar, He, binary mixtures | Hydrogen and inert gas purity analysis |
| Fidas24 | FID Flame Ionization | THC / VOC | Organic carbon measurement; requires hydrogen and combustion air |
| Electrochemical Oxygen Module | Electrochemical | O₂ | Low-cost economical oxygen sensor, for non-environmental CEMS |
Combination rule: One EL3020 / EL3040 chassis can mount max. 2 modules freely combined (例えば. Uras26+Magnos206; Uras26+Limas23; Caldos27+ZO23)
- Sample Complete System Codes (Magnos206 Example for Reference Format)
形式: `24042-XXXXXXXXXX` (system number)
EL3020 + Magnos206, IP20, 2 へ, no flow/pressure sensor: 24042-111011000000
Matching housing: 24342-111011000001; electronic board: 24442-1100110000N1; measuring module: 24641-1110110500H1
EL3040 + Magnos206, IP65, 2 へ, no flow/pressure sensor: 24042-121031000000
Matching housing: 24342-121031000001; electronic board: 24442-1200310000N1; measuring module: 24641-1210310500L1
The prefix for other combinations (Uras26, Limas23, Fidas24 etc.) is identical. Middle digits change with module type, I/O, 電源, 認証 (QAL3) and built-in pump/valve. Formal ordering code must be issued by ABB sales based on final configuration.
- Common Spare Part Order Numbers (Compatible with EL3020 / EL3040)
| 部品番号 | 説明 | Applicable Module |
| 769305 | Modulated light source assembly (dual radiator CaF₂) | Uras26 |
| 769342 | Combined gas capillary assembly (without catalyst) | EL3020 complete unit |
| 9AAC176461 | Fidas24 piping kit | Fidas24 |
| 9AAC176378 | Uras26 modulation unit | Uras26 |
Other consumables: pump diaphragm, FFKM O-ring, filter cartridge, FID nozzle, flow sensor and pressure sensor use separate short codes extracted according to module revision.
- 選択 & Ordering Rules for ABB Inquiry
- シャーシ: EL3020 / EL3040
- モジュール 1 モデル + 範囲 + component; モジュール 2 モデル + 範囲 + component (None if only one module)
- 電源: 100–240 VAC
- I/O: quantity of 4–20 mA AO, ディ/ドゥ
- コミュニケーション: Modbus TCP/RTU, プロフィバス-DP
- Built-in gas circuit: sample pump, solenoid valve bank, filter or not
- 認証: QAL3 (EN14181) for CEMS or not
- 付属品: flow sensor, pressure compensation sensor, calibration valve bank
- Distinguish Easily Confused Models
EL3000 = series collective name
EL3020 = 19-inch rack chassis
EL3040 = wall-mounted IP65 chassis
EL3060 = standalone Magnos206 paramagnetic oxygen analyzer (not EL3020/EL3040 chassis module)
Uras26 / Limas23 / Magnos206 / Fidas24 etc.: measuring modules, not complete analyzer models
- 補足事項
- No public full fixed part number master list, as it is configurable. Any change in range, I/O, certification or gas accessories will alter the trailing code.
- I can provide an EL3000 inquiry template (Excel field list) for you to send to ABB distributors for formal code and quotation.
- Separate spare parts lists for Uras26, Limas23 and Fidas24 are also available on request.
コンタクタ,サーキットブレーカー,センサー,エンコーダ,PLC,コンバータ

















NH42-63-318x560.png)



