APG‑500C 정밀 저항 (피라니) 진공 트랜스미터

성명: APG-500C Precision Resistance (피라니) 진공 트랜스미터 (Pirani Gauge) Commonly known as Chengzhen vacuum transmitter gauge, equivalent to INFICON PSG500 Brand: Chengdu Zhenghua Electronic Instrument, Trademark: 【Chengzhen】 ✅ Key Specifications Measuring Principle: 피라니 (thermal conduction resistance vacuum gauge) 측정 범위: 5×10⁻² Pa ~ 1×10⁵ Pa (공기 / N₂) 전원공급장치: 13.5~30VDC, 소비전력 ...

  • 제품 세부정보

성명: APG-500C Precision Resistance (피라니) 진공 트랜스미터 (Pirani Gauge)

Commonly known as Chengzhen vacuum transmitter gauge, equivalent to INFICON PSG500

상표: Chengdu Zhenghua Electronic Instrument, Trademark: 【Chengzhen】

✅ 주요 사양

측정원리: 피라니 (thermal conduction resistance vacuum gauge)

측정 범위: 5×10⁻² Pa ~ 1×10⁵ Pa (공기 / N₂)

전원공급장치: 13.5~30VDC, Power consumption ≤2W, Max inrush current <300mA

아날로그 출력: Logarithmic voltage 1.9~10.0 VDC (0~0.5V indicates filament failure)

전기적 인터페이스: RJ45 (8-core FCC68)

Vacuum Flange: Standard KF16 (mainstream version)

Max Pressure Withstand: ≤0.2 MPa (absolute pressure)

응답 시간: 100 ms

Mounting Requirement: 수직 설치, keep away from heat sources and vibration sources

적용 가능한 미디어: 공기, N₂, O₂, CO and other inert gases only; Combustible, explosive and strongly corrosive gases are prohibited

📌 RJ45 Pin Definition

핀정의
1힘 + (13.5~30V DC)
2전원 0V (접지)
3아날로그 신호 +
4Transmitter Identification
5아날로그 신호 –
6Relay Contact
7N/C
8Relay Contact

Relay Rating: DC30V / 500엄마 (Only APG500CC supports multi-point threshold control; standard APG-500C has no adjustable control points)

📎 Gas Correction Factor (Reading × C)

Air/N₂/O₂/CO=1.0;H₂=0.5;He=0.8;Ar=1.7;CO₂=0.9;Xe=3.0;Kr=2.4;Ne=1.4;Water vapor=0.5

🛠 Calibration Notes

  1. Preheat for at least 10 min before calibration; continuous power-on is recommended to maintain thermal balance
  2. ATM full-scale calibration: At atmospheric pressure, set to 1×10⁵ Pa → 10.0 VDC
  3. HV zero calibration: Pump down to ≤1×10⁻² Pa, set to 1×10⁻² Pa → 1.5 VDC

📌 Matching Products & 대안

Matching indicator: PDM-5203, PDM-520C

Import equivalent: INFICON PSG500

Upgraded model in same series: APG-500CC (with adjustable relay threshold)

Common failure: Platinum wire contamination, platinum wire burnout (fault output 0~0.5V). Contamination damage is excluded from warranty

📋 일반적인 응용 프로그램

Coating machines, vacuum furnaces, leak detection systems, vacuum packaging, negative pressure processes, coarse & medium vacuum monitoring for semiconductor

APG-500C (Chengzhen/Zhenghua) 문제 해결 체크리스트

전제조건: Distinguish between 【gauge failure】, 【wiring/acquisition side failure】 and 【vacuum system leakage / process issue】 first; 0~0.5V output = filament failure (끊어진 전선 / heavy contamination). This is a hard fault, and contamination damage is not covered by warranty.

Required tools: Multimeter for RJ45 pin voltage measurement, megohmmeter, standard vacuum source, leak detector

  1. 잘못: Fixed output 0~0.5V (fault voltage)

징후: Voltage locked at 0~0.5V, reading unchanged regardless of vacuum, indicator reports fault

근본 원인:

  1. Burned platinum thermal wire (most frequent; easily damaged by repeated atmospheric shock, sudden venting or overpressure)
  2. Heavy oil contamination / coating dust covering the thermal wire inside the gauge, causing thermal conduction failure
  3. Broken wire inside RJ45 plug, poor contact on pin1/pin2 power supply
  4. Damaged drive PCB inside gauge

문제 해결 단계:

① Measure pin1 & pin2 with multimeter to confirm stable 13.5~30VDC supply

② Test with a confirmed good dedicated RJ45 cable

③ If power supply is normal and output remains 0~0.5V after cable replacement → sensor defective

치료: Replace the gauge head directly. Do not disassemble and clean the thermal wire by yourself (easy to cause permanent damage and void warranty)

  1. 잘못: No output at all, no signal on meter, floating reading on PLC

징후: No voltage at signal terminals, same as no connection

근본 원인:

  1. 전원 공급 장치 없음, 역 극성, 끊어진 퓨즈
  2. Broken core inside RJ45 cable, bad crimp on crystal head, broken signal core pin3/pin5
  3. Mismatched input impedance of acquisition terminal (required ≥10kΩ)

문제 해결 단계:

① Measure voltage pin1-pin2, verify polarity (pin1=24V+, pin2=GND)

② Power the gauge separately and measure analog output directly at pin3 & pin5 to rule out rear-end wiring/module faults

③ Confirm PLC/display meter supports high-impedance voltage input; do not use current input channel

치료: 케이블 교체, correct power polarity, use compatible acquisition channel

  1. 잘못: Inaccurate reading, large deviation, wrong value at atmosphere or high vacuum

징후: Not 1E5Pa at atmospheric pressure, cannot reach expected vacuum after pumping, large deviation compared with reference gauge

근본 원인:

  1. No preheating or calibration (≥10min preheating required)
  2. Gas is not air/N₂ and gas correction factor not applied (large deviation for Ar, He, H₂)
  3. Slight contamination or long-term drift of thermal wire
  4. Wrong mounting orientation (non-vertical installation, condensate/dust accumulates on thermal wire)
  5. Leakage at KF flange, aged KF O-ring, misaligned gasket
  6. Gauge inlet partially blocked by oil mist or dust

문제 해결 단계:

① Atmosphere calibration: Vent to atmosphere, stabilize for 10 분, perform atmospheric calibration

② Pump down to ≤1×10⁻²Pa for zero calibration

③ Confirm gas type and apply correction factor

④ Check mounting posture: 수직 설치, flange downward or lateral; avoid upside-down installation which causes liquid accumulation

⑤ Isolate gauge and perform leak test for chamber leak rate

치료: Recalibrate, adjust installation, replace sealing ring, add front oil baffle; replace sensor if heavily contaminated

  1. 잘못: Reading fluctuates continuously, severe drift and instability

징후: Vacuum is stable but reading jumps back and forth

근본 원인:

  1. Large ripple / unstable power supply, shared power supply with high-power motors or inverters
  2. Signal cable routed together with power cable, 전자기 간섭
  3. Intrinsic pressure fluctuation of system (pump pulsation, frequent valve switching, airflow disturbance)
  4. Slight contamination or aging of thermal wire
  5. 접지 불량, mixed analog signal ground and power ground

문제 해결 단계:

① Use isolated DC power supply for independent gauge power supply

② Single-end grounding for cable shield, keep away from power cables

③ Remove gauge and power on in static atmosphere; if stable at atmosphere → fault from system airflow / 누출

치료: Add shielding, isolated power supply, optimize grounding; add damping tube on chamber

  1. 잘못: Permanent atmospheric reading (~10V), reading does not drop while pumping

징후: Pump keeps running, gauge stays at atmospheric pressure reading

근본 원인:

  1. Severe leakage at KF flange, missing or damaged O-ring
  2. Isolation valve before gauge closed, gauge separated from chamber
  3. Gauge inlet fully blocked by solid deposit, gas cannot enter gauge cavity
  4. 내부 회로 결함, output locked at full scale

문제 해결 단계:

① Close gauge isolation valve, test gauge under vacuum separately to distinguish gauge fault or system leakage

② Visual check KF O-ring and clamp compression

③ Disassemble flange and check inlet for coating material / dust blockage

치료: Leak detection, open valve, clean inlet; replace sensor if blocked and thermal wire contaminated

  1. 잘못: Abnormal relay setpoint (APG-500CC relay version only)

징후: Relay does not trigger / false trigger at target vacuum

근본 원인:

  1. Wrong threshold calibration procedure (incorrect timing of setting button)
  2. Relay contact overloaded (max DC30V/500mA; cannot drive high-power load directly)
  3. Oxidized or welded contacts
  4. 신호 드리프트, mismatch between actual vacuum and gauge output voltage

문제 해결 단계:

① Relearn threshold following manual (hold setting pin, reach target vacuum then release)

② Test contact continuity with multimeter under no-load condition

③ Add intermediate relay for load isolation

치료: Recalibrate; replace whole transmitter if contacts damaged

  1. 잘못: Frequent filament burnout (high failure rate)

Trigger list (engineering prevention focus)

  1. Fast frequent venting of chamber, instantaneous airflow impact on thermal wire
  2. Continuous deposition of process oil mist and evaporation coating material on thermal wire
  3. Overpressure shock (>0.2 MPa absolute)
  4. Combustible / corrosive gas entering gauge tube (prohibited for this model)
  5. Upside-down installation, condensate dripping directly onto thermal wire

대책: Install front oil trap, slow venting, add isolation valve, 수직 설치, restrict process gas

✅ 빠른 자가 점검 순서 (on-site recommended order)

  1. 공급 전압 측정 (pin1, pin2) → rule out reverse polarity / 저전압
  2. Measure analog output (pin3, pin5), first check if it is 0~0.5V fault state
  3. Cross-test with spare cable to separate gauge fault or cable / backend acquisition fault
  4. Static atmosphere test, check whether reading stabilizes at atmospheric pressure
  5. Isolated vacuum test to distinguish 【gauge drift】 or 【chamber leakage / pump issue】
  6. 구경 측정, check mounting posture, oil baffle and sealing

⚠️ Important Disclaimer

Thermal wire contamination, oil fouling and burnout caused by dust are not covered under warranty. The manufacturer usually only supplies replacement gauge heads instead of repair.

Do not disassemble gauge head or flush internal thermal wire with solvent, permanent damage may occur immediately.

Pirani gauge is gas-dependent and not universal vacuum gauge; correction is mandatory for Ar, He, H₂.

Chengzhen (Chengdu Zhenghua ZHAC) 전체 모델 목록

상표: Chengzhen / Chengdu Zhenghua Electronic Instrument; grouped by category; APG Pirani transmitter (your main model) 상단에

  1. 피라니 / Resistance Vacuum Transmitters (Integrated transmitter gauge, PLC direct acquisition, equivalent to INFICON PSG500)

APG-500C: Standard Pirani resistance transmitter, RJ45, logarithmic output 1.9~10V, KF16

APG-500CC: Version with adjustable relay threshold output

FPG550: 저항 + piezoresistive composite transmitter (넓은 범위, atmosphere down to rough vacuum)

ATG-530: 1Torr range precision transmitter

  1. Vacuum Gauge Tubes (Separate type, requires matching indicator)

Resistance Gauge (피라니)

ZJ-52T: Classic resistance vacuum gauge tube (glass gauge, KF, CF flange options)

Thermocouple Gauge

ZJ-51

ZJ-53B

ZJ-54D (Short-filament thermocouple, widely used in coating industry)

Piezoresistive Gauge

ZJ-1P: Piezoresistive vacuum gauge, 0.1~100kPa, suitable for near-atmosphere rough vacuum

Capacitance Diaphragm Gauge (Gas-independent, 높은 정확도)

ZJ-1C Series (Range options: 10티, 100티, 1000티, 0~1000Pa; 4-20엄마 / 0-10V 출력)

Hot Cathode Ion Gauge (B-A gauge, high vacuum)

ZJ-2

ZJ-10

ZJ-12 (B-A ultra-high vacuum gauge)

ZJ-27 (Wide-range ion gauge, mainstream for coating machines)

Cold Cathode Ion Gauge (Penning / inverted magnetron cold gauge, high vacuum, no filament)

ZJ-14 (Inverted magnetron cold gauge)

ZJ-14C (Penning cold gauge)

  1. Single-type Vacuum Indicators (Panel mount for control cabinet)

ZDR Resistance (피라니) Vacuum Meter

ZDR-I-Pro: Single-channel resistance vacuum meter

ZDR-II-Pro: Dual-channel resistance vacuum meter

ZDR-X-LED: Rack-mount multi-channel resistance vacuum mete

ZDO Thermocouple Vacuum Meter

ZDO-I / ZDO-II: 하나의 / dual channel thermocouple vacuum meter

ZDK Hot Cathode Ion Vacuum Meter

ZDK-I-LED, ZDK-X-LED: 하나의 / multi-channel ion meter, compatible with ZJ-27/ZJ-

ZDV Cold Cathode Ion Vacuum Meter

ZDV-I-LED, ZDV-IC-LED: Dedicated indicator for cold gauge

PDM Miniature Resistance Vacuum Meter (콤팩트, portable all-in-one)

PDM-5203: Panel miniature resistance vacuum meter (match APG, ZJ-52T)

PDM-520B: Handheld miniature vacuum meter (5V 전원, field portable)

  1. ZDF Composite Vacuum Meter (업계 주류, 저항 + ion / 열전대 / cold gauge combination, standard for coating machines)

ZDF-I-LED / ZDF-I-PRO: 열전대 + ion composite

ZDF-III-LED / ZDF-III-PRO: 저항 (ZJ52T) + hot cathode ion (ZJ27), 가장 일반적인

ZDF-IV-LED / ZDF-IV-PRO: 저항 + B-A ultra-high ion gauge (ZJ12)

ZDF-V-LED / ZDF-V-PRO: 저항 + cold cathode cold gauge (ZJ14)

ZDF-X-LED / ZDF-X-PRO: Dual resistance + single ion

ZDF-IB / ZDF-VB / ZDF-VC: Dual thermocouple, dual resistance cold gauge, dual ion and other special composite versions

  1. Automatic Pressure Controller (Vacuum closed-loop pressure stabilization control)

ZDC-I: 저항 + ion composite pressure controller

ZDRC-I: Resistance pressure controller

ZDMC-I: Capacitance gauge pressure controller

ZDFC-III: Composite vacuum pressure controller

  1. 어울리는 액세서리

ZDT-5: Electromagnetic fine-tuning vacuum valve (for pressure controller)

Ceramic sealed vacuum feedthrough electrode (vacuum electrode, vacuum feed wire)

KF/CF flange adapter, gauge cable, dedicated APG RJ45 cable

Quick Selection Reference for BOM

Rough vacuum, PLC acquisition, replace PSG500 → APG-500C / APG-500CC

Indicator display + rough vacuum → ZDR series + ZJ-52T

Standard coating machine composite vacuum (1e5~1e-5Pa) → ZDF-III-PRO + ZJ52T+ZJ27

Ultra-high vacuum → ZDF-IV + ZJ12

Filament-free, long-term high vacuum → ZDF-V + ZJ14C

Gas-independent high-accuracy absolute pressure → ZJ-1C capacitance gauge