흡충 572-2 High Temperature Infrared Thermometer - 접촉기,회로 차단기,감지기,인코더,PLC,변환기

위챗: +86-13184948252 왓츠앱: 0086-13811255435 이메일: kent@bestcontactor.com

에 대한 연락하다 |

 흡충 572-2 High Temperature Infrared Thermometer - 접촉기,회로 차단기,감지기,인코더,PLC,변환기

전기 같은/

흡충 572-2 High Temperature Infrared Thermometer

제품 포지셔닝: Industrial handheld infrared thermometer combining infrared and K-type thermocouple. Designed for long-distance high-temperature measurement, widely used in power, 야금, cement and petrochemical industries. Key Specifications Infrared Temperature Range: -30 ℃ ~ 900 ℃ Distance-to-Spot Ratio (디:에스): 60:1 (90% 에너지), minimum spot diameter φ19 mm, dual-laser targeting Infrared Accuracy (at 23℃ ambient): ≥0℃: ±1℃ ...

  • 제품 세부정보

제품 포지셔닝: Industrial handheld infrared thermometer combining infrared and K-type thermocouple. Designed for long-distance high-temperature measurement, widely used in power, 야금, cement and petrochemical industries.

주요 사양

Infrared Temperature Range: -30 ℃ ~ 900 ℃

Distance-to-Spot Ratio (디:에스): 60:1 (90% 에너지), minimum spot diameter φ19 mm, dual-laser targeting

Infrared Accuracy (at 23℃ ambient):

≥0℃: ±1℃ or ±1% of reading, whichever is greater

-10 ~ 0℃: ±2℃

<-10℃: ±3℃

K-type thermocouple input (contact temperature measurement): -270 ℃ ~ 1372 ℃

Emissivity: 0.10 ~ 1.00 조절할 수 있는 (0.01 step), built-in material emissivity lookup table

응답 시간: <500ms (95%)

Spectral Band: 8~14 μm

데이터 저장: 최대 99 기록, USB export supported

표시하다: MAX/MIN/AVG/DIF, high/low temperature audible & 시각적 경보, adjustable backlight

전원공급장치: 2×AA batteries; 순중량: 322g

주요 특징

  1. Dual-Laser Targeting: Offset dual lasers mark the measurement area for precise targeting of small objects from long distances
  2. Dual Measurement Mode: Non-contact infrared + K-type thermocouple contact measurement, balancing operational safety and accuracy verification
  3. Industrial Robustness: Ergonomic anti-slip housing, intuitive menu, 다국어 지원
  4. Alarm Function: Audible and visual alert on temperature threshold violation, suitable for inspection and predictive maintenance of equipment

일반적인 응용 분야

전력산업: Switchgear, 모선, 변압기, cable joints temperature inspection

야금 / Heat treatment: Furnace body, rolls, casting surface temperature

시멘트, 유리, 석유화학: Kiln, pipeline, reactor surface temperature measurement

Machinery industry: 모터, bearing and high-temperature equipment inspection

표준 패키지 내용물

흡충 572-2 Main Unit

K-type bead thermocouple probe

USB cable

2 AA batteries

User manual and certificate of conformity

Comparison with Related Models

흡충 572-2: 60:1 비율, max infrared temp 900℃, with K-type thermocouple and data storage

흡충 568: 50:1 비율, max infrared temp 800℃

흡충 62 MAX+: 12:1 비율, max temp 650℃, entry-level industrial model

선택 & 작동 참고 사항

  1. 디:S=60:1: The spot size increases with measurement distance; shorten distance when measuring small targets
  2. Polished metallic surfaces have low emissivity and may cause large measurement error. High-temperature tape is recommended to raise emissivity
  3. Infrared measures surface temperature only; K-type thermocouple is used for contact temperature comparison
  4. Not suitable for measurement through glass or transparent plastics (infrared radiation will be blocked)

흡충 572-2 Troubleshooting Checklist

Recommended troubleshooting sequence: Power supply → Lens / Optical path → Parameter setting → Ambient condition → Hardware / Error code; Do NOT disassemble the unit. Disassembly of internal laser and sensor will void calibration and warranty.

  1. Power Supply Faults
징후근본 원인치료
Blank screen, no response when trigger is pulledSleep mode / Dead battery / 역극성1. Pull trigger to wake up; 2. Replace with new AA alkaline batteries, verify polarity; 3. Inspect battery contacts for oxidation or electrolyte corrosion
Low battery icon displayed, unstable readingsInsufficient battery voltageReplace batteries immediately; Accuracy is unreliable under low power, discontinue inspection
Emissivity, time and stored data lost after battery replacementBattery removal exceeds 1 분, mainboard power resetReplace battery promptly upon low power alert; Complete battery swap within 1 분; Reconfigure parameters afterwards
  1. 표시하다 & Error Codes
징후근본 원인치료
shown on screenTarget temperature out of range (<-30℃ or>900℃)Confirm target temperature is within measuring range; Infrared max 900℃, K-type thermocouple max 1372℃
EEPROMErr / CalAreaErr / ProbCalErDamaged internal storage / calibration dataNo on-site repair possible. Stop operation and send to Fluke authorized service center
Alarm light flashes with buzzer sound높은 / low temperature alarm triggeredEnter menu to check threshold, modify or disable alarm as required

III. Laser Targeting Faults

징후근본 원인치료
Valid infrared reading but dual laser does not turn onLow battery; Ambient temp>40℃; Laser disabled in menu1. Replace battery; 2. Move to location with ambient temp ≤40℃; 3. Enable laser in menu
Blurred or offset laser spotDirty or scratched lens렌즈 청소; Replace lens assembly if scratched
  1. Inaccurate / Drifting Readings (Most Common Fault)

Distinguish first: Infrared reading error VS K-type thermocouple probe error

Inaccurate Infrared Readings

  1. Wrong emissivity setting: Polished metal has low emissivity leading to high reading; Non-metal has high emissivity. Adjust emissivity from 0.10~1.00; Apply high-temperature tape for shiny metal (emissivity≈0.95)
  2. Field-of-view spot issue: 디:S=60:1. Too long distance results in spot larger than workpiece and captures surrounding hot/cold objects. Shorten distance and ensure full spot coverage on target surface
  3. Lens contamination: 먼지, 기름, moisture attenuate infrared signal

✅ Cleaning procedure: Blow with compressed air → Brush gently with camel hair brush → Wipe with cotton swab moistened with pure water; Do NOT use alcohol, acetone or other solvents, which will damage infrared lens

  1. Ambient interference: Hot air stream, 증기, smoke, 유리 / transparent plastic blocking optical path; 강한 전자기 간섭 (용접기, 유도 가열); water film or oxide scale on target surface
  2. Thermal shock: Sensor unstable when moved from drastically different temperature environment. Let instrument stabilize ≥30 minutes on site before measurement
  3. Reflected heat from surrounding hot sources causes falsely high reading: Adjust measurement angle to avoid hot background reflection

No Reading / Abnormal Reading on K-type Thermocouple Channel

  1. Loose or oxidized probe connector; 끊어진 전선

✅ Remedy: Re-seat connector; Cross-test with 80PK series K-type probe to identify fault on probe or instrument jack

  1. Wrong thermocouple type selected: Unit only accepts K-type. S/T/J type probes cannot read correctly
  2. Cold junction compensation anomaly: Large temperature difference at probe terminal
  3. USB Communication & Data Export Faults
징후근본 원인치료
PC fails to recognize instrument and export recordsInstrument not powered on; FlukeView Forms software not running; Damaged USB cable1. Hold trigger to keep unit powered; 2. Launch FlukeView Forms software; 3. Replace USB cable for test; 4. Check PC USB port
Stored data lostBattery depleted for long time; Battery replacement timeoutExport records regularly; Replace battery at low power prompt
  1. 기계 & Appearance Faults
  2. Sticky trigger: Dust accumulated in trigger gap, clean with dry brush, Do NOT apply lubricant
  3. Cracked housing: Damage from drop impact. Crack does not affect temperature reading but reduces ingress protection, keep away from moisture
  4. Lens scratch: Scratch introduces measurement error, return to factory for lens replacement and recalibration

Ⅶ. Fault Judgment & Service Standard (When to Stop Using)

✅ On-site repairable: Battery replacement, lens cleaning, parameter reset, probe cross-verification

❌ Must send to official service, no disassembly allowed:

  1. EEPROM or calibration related error codes
  2. Cracked lens, internal water condensation
  3. Persistent reading deviation after drop impact, cannot be corrected by lens / emissivity adjustment
  4. Battery leakage corroding mainboard contacts

Ⅷ. Quick On-Site Self-Check Procedure

  1. Install new batteries and stabilize for 30 min under ambient condition
  2. 렌즈 청소, restore factory menu setting, set emissivity to 0.95
  3. Measure warm water (대략. 80~90℃) in infrared mode, connect K-type thermocouple for contact comparison; Deviation shall stay within specified accuracy
  4. Verify laser, high/low alarm, backlight and trigger hold function
  5. Test USB connection and data read/write

흡충 570 시리즈 (572 Family) 모델 목록

Consists of Legacy 570 시리즈 (단종) and Currently available 572-2. 흡충 570 series is industrial high-precision long-distance infrared thermometer with K-type thermocouple input.

  1. Legacy 570 시리즈 (단종, available in spare / second-hand market)
모델Key Configuration주요 차이점
흡충 572 (유산, no -2 접미사)-30~900℃, 디:S=60:1, 3-point laser targeting, K-type thermocouple input, RS232 interfaceOld version with serial port, no USB, 단종
흡충 574-30~900℃, 디:S=60:1, 3-point laser, RS232, 데이터 로깅, optional close-focus 50:1 작은 점Enhanced data logging vs 572; 574-NI: version without infrared calibration certificate
흡충 576-30~900℃, 디:S=60:1, built-in digital camera for image archive, USB, 3-point laserTop-tier model with image capture, long discontinued
  1. Current Active Model: 흡충 572-2 (Replaces legacy 572)

Your target model, current main production unit

온도 측정: 적외선 -30 ~ 900℃; K-type thermocouple -270 ~ 1372℃

디:S=60:1, dual-laser targeting; 99 local storage records, USB export

Dual mode (적외선 + contact thermocouple), 들리는 & visual high/low temperature alarm

산업: 힘, 야금, 석유화학, cement high-temperature inspection

III. Related Dual-Mode Infrared Thermometers (Commonly compared with 572-2 for selection)

Not part of 570 시리즈, but Fluke handheld infrared + K-type thermocouple models frequently referenced in project selection

|모델|Infrared Range|디:에스|특징|

|—|—|—|—|

|흡충 568|-30~800℃|50:1|Dual mode, lower spec alternative to 572-2 with lower max temperature|

|흡충 568 EX|-30~800℃|50:1|Intrinsically safe explosion-proof version for hazardous oil & gas area|

|흡충 561|-40~550℃|12:1|HVAC dedicated entry-level dual mode unit|

  1. Infrared-only Models (No thermocouple, reference for category comparison, not in 570 시리즈)

흡충 62 MAX / 62 MAX+: -30~650℃, 디:S=12:1, IP54, 3m drop resistance

흡충 64 MAX: -30~600℃, 디:S=20:1

  1. Model Version Remarks
  2. 572-2 has no sub-models, only standard main unit. Variants come from accessory kits: Optional carrying case, extended thermocouple probe, tripod
  3. Legacy 572/574/576 permanently discontinued. Only surplus stock, refurbished or second-hand units available; calibration and spare parts are difficult to source. 선택하다 572-2 for new projects
  4. Key distinction: Legacy models use 3-point laser; 새로운 572-2 uses dual laser targeting system