Catalog No.: 9385100 (CN Chinese Version)–
مبدأ: LED colorimetric method, used for rapid on-site water quality testing with Hach preformed reagents
✅ المواصفات الأساسية
مصدر الضوء: قاد, 4 fixed wavelengths: –420 / 520 / 560 / 610 نانومتر–, wavelength accuracy ±1 nm
الكاشف: Silicon photodiode
Photometric range: 0~2 Abs; photometric accuracy ±0.03 Abs
Readout modes: Concentration, Absorbance (Abs), % Transmittance
Pre-programmed methods: –90 built-in methods–; يدعم ما يصل إلى 10 custom user curves
Data storage: 500 GLP-compliant records (وقت, sample ID, user ID), export via Mini USB
Sample cell: 1-بوصة (25 مم) round sample vial; متوافق مع 16 mm digestion tubes (COD tubes) via adapter
حماية الدخول: –IP67 (when the cell lid is closed)–, shock-resistant and drop-proof
مزود الطاقة: 4 × بطاريات AA; backlit graphic display (240×160), supports Chinese plus 21 other languages
أبعاد: 231 × 96 × 48 مم; الوزن تقريبا. 0.6 كجم
درجة حرارة التشغيل: 10~40 ℃, humidity ≤80% non-condensing
✅ Typical Measurable Parameters
Free chlorine, total chlorine, ammonia nitrogen, COD, total phosphorus, total nitrogen, الكروم سداسي التكافؤ, total chromium, نحاس, حديد, manganese, zinc, fluoride, التعكر, الألومنيوم, barium, إلخ.
Applicable scenarios: Municipal wastewater, ماء الصنبور, circulating cooling water, industrial wastewater, surface water, aquaculture, إلخ.
✅ الملحقات القياسية
DR900 main unit, 2 × 25 mm glass sample vials, 2 × 1 cm plastic cuvettes, 16 mm COD tube adapter, 4 × بطاريات AA, Mini USB cable, multilingual manual, CD documentation
✅ الميزات الرئيسية
- Portable field unit, IP67 dust and water resistance, ideal for field work and wastewater plant inspections
- Favorites function for one-touch access to frequently used methods
- Supports blank correction and standard correction to offset reagent batch variations
- Data export to PC for environmental audit and record-keeping compliance
- Designed exclusively for Hach pre-packaged powder pillows and Test’N Tube digestion reagents; not suitable for arbitrary use with self-prepared reagents
✅ Comparison Reference
DR900: –4 fixed-wavelength colorimeter–, cost-effective for rapid field testing; no spectral scanning capability
DR1900: Portable spectrophotometer with continuously adjustable wavelength, suitable for both laboratory and field use
✅ Common Spare Parts & المواد الاستهلاكية
Sample vials, cuvettes, 16 mm adapter, batteries, dust cover, matched preformed reagents (COD, ammonia nitrogen, total phosphorus, إلخ.)
HACH DR900|Common Parameters – Method Numbers – Range List (High-frequency for Municipal Wastewater / Tap Water / تعميم المياه)
ملحوظة: –Different ranges for the same parameter correspond to different method numbers; the matching preformed reagent must be used.Unit: ملغم/لتر; chlorine expressed as Cl₂; ★ denotes EPA-approved methods for environmental reporting
| المعلمة | Method No. | يتراوح | Reagent Format | ملاحظات |
| COD (Low Range LR) | 8000 | 0.7~40 | TNT Tube | Surface water, low‑concentration wastewater |
| COD (Mid Range MR) | 8000 | 3~150 | TNT Tube | ★ Most common for municipal wastewater |
| COD (High Range HR) | 8000 | 20~1500 | TNT Tube | Industrial wastewater |
| COD (Ultra‑high Range) | 8000 | 200~15000 | TNT Tube | High‑strength wastewater |
| Ammonia Nitrogen (Nessler LR) | 8038 | 0.02~2.50 | Powder Pillow | Tap water, surface water |
| Ammonia Nitrogen (Nessler HR) | 8038 | 0.1~10.0 | Powder Pillow | Wastewater plant effluent |
| Ammonia Nitrogen (Salicylate Method) | 10031 | 0.01~0.50 | Powder Pillow | Low concentration, less interference |
| Total Phosphorus (Ammonium Molybdate LR) | 8190 | 0.02~2.50 | Powder Pillow | مياه الصرف الصحي |
| Total Phosphorus (TNT) | 10209 | 0.06~3.50 | TNT Tube | Digestion required |
| Orthophosphate | 8048 | 0.02~2.50 | Powder Pillow | No digestion required |
| Total Nitrogen | 10071 | 1~16 | TNT Tube | Persulfate digestion, ★ |
| Free Chlorine (DPD) | 8021 | 0.02~2.00 | Powder Pillow | Tap water inspection |
| Free Chlorine (DPD HR) | 10069 | 0.1~10.0 | Powder Pillow | High‑chlorine disinfection water |
| Total Chlorine (DPD) | 8167 | 0.02~2.00 | Powder Pillow | Circulating water |
| Total Chlorine (DPD HR) | 10070 | 0.1~10.0 | Powder Pillow | High chlorine |
| Chlorine Dioxide | 10126 | 0.04~5.00 | Powder Pillow | Chlorine dioxide disinfection |
| Hexavalent Chromium Cr⁶⁺ | 8023 | 0.01~0.60 | Powder Pillow | Industrial wastewater |
| Total Chromium | 8024 | 0.01~0.60 | Powder Pillow | Digestion required |
| Total Iron (FerroVer) | 8008 | 0.02~3.00 | Powder Pillow | Circulating water, boiler water |
| Ferrous Iron | 8147 | 0.02~3.00 | Powder Pillow | No digestion required |
| Manganese | 8149 | 0.01~2.00 | Powder Pillow | Groundwater |
| نحاس | 8506 | 0.04~5.00 | Powder Pillow | Industrial water |
| الزنك | 8009 | 0.05~3.00 | Powder Pillow | Electroplating wastewater |
| Fluoride | 8029 | 0.1~2.0 | SPADNS Reagent | Surface water |
| Total Hardness (as CaCO₃) | 8030 | 2~1000 | Powder Pillow | Circulating water |
| Turbidity | 8660 | 0~1000 NTU | القياس المباشر | No reagent required, sample vial only |
| الألومنيوم | 8012 | 0.008~0.80 | Powder Pillow | Aluminum residual in tap water |
| Barium | 10251 | 5~100 | Powder Pillow | Oilfield produced water |
| Bromine | 8016 | 0.05~4.50 | Powder Pillow | Bromine disinfection |
| Phenol | 8044 | 0.002~0.12 | Powder Pillow | Volatile phenol |
| TOC | 10129 | 0.5~25 | TNT Tube | Total Organic Carbon |
Quick Operating Notes
- For one method number (على سبيل المثال. 8000 for COD), –range differentiation is identified by reagent tube barcode / يتراوح–. Select the same method number on the instrument and change reagents for different ranges.
- TNT digestion tubes must be used with the 16 mm adapter in DR900; powder pillow tests generally use the 25 mm round sample vial.
- Turbidity test 8660: Fill vial directly with sample, use pure water for blank; no reagent needed.
- Digestion-required parameters (total nitrogen, TNT total phosphorus, total chromium): DRB200 / DRB900 digester required.
- 90 methods are preloaded in total; the table above covers more than 30 commonly used engineering methods. Niche methods for boron, hydrazine, sulfide, إلخ. can be extracted separately.
HACH DR900 Troubleshooting Checklist
Preface: The DR900 is a fixed 4-wavelength LED colorimeter. Most faults originate from –optical path, sample vial, adapter, مزود الطاقة, data and operation–. Diagnose by error code / symptom and troubleshoot from simple to complex.
- Power-on Faults
-1. لن يتم تشغيله / Black screen on startup
بطارية منخفضة: Replace with fresh AA alkaline batteries; carbon batteries are not recommended due to unstable voltage
Poor battery contact: Remove batteries, clean metal contacts in battery compartment, check for bent terminals
Battery polarity reversed: Match +/marking inside compartment
Corrosion from water ingress in battery compartment: Wipe stains and green corrosion with anhydrous alcohol and fully dry before retest; heavy corrosion requires factory repair
Power note: USB connection is for data export –فقط; cannot power the unit for startup–
-2. Restarts repeatedly / Shuts down automatically after power-on
Aged high-internal-resistance batteries: انخفاض الجهد تحت الحمل; retest with new batteries
Ambient temperature out of specification: <10 ℃ or >40 درجه مئوية; battery capacity drops sharply at low temperature
Loose battery contacts: Gently pry contact tabs outward to improve contact pressure
- Optical Path & Reading Abnormalities (الأكثر تكرارا)
-3. Unstable readings, continuous fluctuation, heavy drift
Fingerprints, water stains or reagent residue on outer vial surface: Wipe outer wall with lint-free soft cloth; only grip the frosted zone
Scratched or worn sample vial: Scattering from scratches; replace vial directly
Air bubbles in sample: Allow to settle; avoid vigorous shaking. Tap TNT COD tubes gently to release bubbles
Turbid sample or sediment: Let sample settle or wait for full color development per reagent instructions
Misaligned vial seating: Align vial with slot marker and notch with positioning feature
Light source aging: Abnormally high blank absorbance, poor repeatability on identical sample; suspect LED degradation
Direct strong ambient light: Shield sample cell from sunlight, which interferes with optical measurement
-4. Blank calibration failure / Blank Fail error
Impure blank water: Use ammonia-free / COD-free pure water; tap water or mineral water cannot be used as blank
Contaminated blank vial: Residual carryover from high-concentration prior tests; soak in acid or replace
Mismatched paired vials for blank and sample: Hach factory vials are recommended to be used as matched pairs
Spilled liquid inside cell / dirty optical window: Open sample compartment, clean quartz window with lint-free cloth dampened with pure water and dry fully
Wrong method selected: Mismatched method and reagent will cause blank failure
-5. Negative readings
Blank vial and sample vial swapped
Sample concentration below method detection limit
Dirty / scratched vial making blank more opaque than sample
Expired / degraded reagent with no color formation
Drifted calibration curve: Perform blank recalibration
-6. Over-range reading (رأ / Over Range)
Sample concentration exceeds upper limit of selected method: Dilute sample manually, select higher-range method and reagent
Wrong reagent dosage or reversed addition sequence
Incomplete digestion for COD / total nitrogen: Insufficient digestion temperature or hold time
ملحوظة: DR900 has no automatic dilution function; manual dilution and retest required
-7. Optical error / Lamp Error
Sample lid not fully closed, poor magnetic switch contact
Optical window covered with liquid or crystals → clean window
Failed LED light source: If error persists after cleaning, battery replacement and lid closure, return for LED assembly replacement
Water ingress in compartment and damp optical PCB: Discontinue use, dry for 24 hours and retest
- TNT Digestion Tube & 16 mm Adapter Related Errors
-8. Inaccurate or erratic COD / total nitrogen results with TNT tubes
Adapter not fully seated or tilted, causing inclined tube
Condensation droplets or reagent crystals on outer wall after digestion: Wipe completely
Tube label / graduation mark blocking optical path: Label must be kept away from light path
Sediment settled across light path: Invert gently and let sediment settle below optical path
Debris or broken glass inside adapter cavity: Clean adapter interior
Never insert a 16 mm digestion tube directly into the 25 mm cell without the adapter
- عملية, طريقة & Reagent Issues
-9. Large deviation between DR900 results and laboratory reference
- Verify three items: –method number, reagent batch, reaction hold time(DPD chlorine fades over time; Nessler ammonia nitrogen drifts with extended standing)
- Check interferences: Sulfide, heavy metals, oxidants and high salinity cause interference (refer to Hach reagent method interference sheets)
- Digestion tests: Confirm DRB200 / DRB900 digester temperature and timing accuracy
- Expired or moisture-damaged reagents: Caked powder pillows, discolored TNT tubes must be discarded
- Sample holding time exceeded: Unpreserved analytes degrade rapidly (especially chlorine and ammonia nitrogen)
-10. Required test method missing
Confirm firmware version; older firmware contains fewer built-in methods. Contact Hach for firmware upgrade support
The DR900 has a fixed library of 90 embedded methods. –Spectral scan methods cannot be added freely–, only up to 10 custom user curves are supported
Lost custom curves: User curves are stored locally; factory reset deletes all custom curves
- Data & USB Communication Faults
-11. USB connection failure, cannot export data
Must use Mini USB data cable; many charging cables lack data conductors. Use original Hach cable
Missing PC driver: Hach uses standard HID class, no special driver required. Better compatibility with Win10/11; some stripped system builds fail recognition
Instrument must be powered on when connected via USB
500-record storage full: Export records then delete old entries. GLP records support single deletion only, no bulk erase
Export software: Hach DataCom; generic serial terminal software is not compatible
-12. Lost records / Incorrect clock time
Depleted internal clock battery (onboard coin cell): Time resets after power-off; return for clock battery replacement and time setup
Factory reset executed accidentally, clearing custom curves and stored records
- الأجهزة, Ingress & Water Damage
-13. Liquid spillage / water entry in sample compartment
IP67 rating applies –only when lid is fully closed–. No water protection when lid is open
Immediately remove batteries after water ingress, leave lid open to air dry ≥24 hours in cool shaded area. Do NOT apply hot air (damages LED and PCB)
If short circuit or corrosion signs are present, avoid repeated power cycles and send for service
-14. Abnormal screen, عرض مشوه, missing segments
Slow LCD response at low temperature: Normal behaviour; restore after warming
Physical impact and casing deformation: Loose ribbon cable or damaged display; return for repair
- قائمة طعام & معايرة
-15. Instrument prompts for standard calibration
Routine DR900 operation only requires –blank measurement per test–. Standard calibration is only needed for custom user curves
Factory embedded methods normally do not require standard recalibration. If persistent bias occurs, verify with Hach certified check standard
- On-site Rapid Troubleshooting Sequence (Field Emergency)
- Install new batteries → 2. Clean sample vials and optical window → 3. Re-run blank calibration → 4. Check sample for bubbles and sediment → 5. Confirm method number, reagent and reaction time → 6. Inspect lid magnet and adapter seating → 7. Check temperature and strong light exposure → 8. Verify with standard sample to isolate instrument vs sample/reagent error
- العمليات المحظورة
Do not clean optical windows with acetone or aggressive solvents (coating damage)
Do not scrub vials with abrasive paper or hard material (permanent scratches)
Do not insert hot / pressurized TNT tubes; cool fully to ambient temperature before measurement
Do not disassemble the housing; disassembly voids warranty
HACH Colorimeter / Portable Spectrophotometer Model & Catalog List (Mainstream China SKUs)
امتياز: –Colorimeter (fixed LED wavelength)مقابل –Spectrophotometer (continuously tunable wavelength)–; PCII is legacy, DR300 is PCII replacement; DR900 multi-parameter fixed wavelength; DR1900 and above are spectrophotometers often grouped together for quotation
ملحوظة: CN = Chinese version; catalog numbers are standard Hach China SKUs; OEM part numbers available for foreign trade
- DR900 Multi-parameter Portable Colorimeter (4 fixed wavelengths)
| نموذج | Catalog No. | وصف |
| DR900 (Standard Chinese Version) | 9385100 | الوحدة الرئيسية + 25 mm vials, 1 cm cuvettes, 16 mm COD adapter, batteries, كابل يو اس بي, يدوي, 90 built‑in methods, IP67 |
- DR300 Pocket Single-parameter Colorimeter (PCII successor, IP67, Bluetooth optional)
| نموذج | Catalog No. | Measured Parameter |
| DR300 | LPV445.80.00110 | Free & Total Chlorine (DPD) |
| DR300 | LPV445.97.01110 | Bromine Br₂ |
| DR300 | LPV445.97.02110 | Nitrate |
| DR300 | LPV445.97.03110 | Dissolved Oxygen |
| DR300 | LPV445.97.04110 | Ozone |
| DR300 | LPV445.97.06110 | Orthophosphate |
| DR300 | LPV445.97.09110 | الزنك |
| DR300 | LPV445.97.10110 | Molybdenum |
| DR300 | LPV445.97.25110 | الألومنيوم |
| DR300 | LPV445.97.26110 | Monochloramine / Free Ammonia |
| DR300 | LPV445.97.40110 | Ammonia Nitrogen |
| DR300 | LPV445.80.51110 | Chlorine Dioxide |
| DR300 | LPV445.97.62110 | Free / Total Chlorine Mid‑High Range |
| DR300 | LPV445.97.22110 | Total Iron (FerroVer) |
| DR300 | LPV445.97.16110 | Iron TPTZ |
- PCII (Pocket Colorimeter II, legacy single-parameter, phased out for DR300)
| نموذج | Catalog No. | المعلمة |
| PCII | 5870000 | Free / Total Chlorine DPD |
| PCII | 5870050 | Chlorine Dioxide |
| PCII | 5870040 | Ozone |
| PCII | 5870030 | Bromine |
| PCII | 5870020 | Iron |
| PCII | 5870010 | Manganese |
- SL Series Portable Parallel Analyzer (PPA, multi-channel rapid field)
| نموذج | Catalog No. | وصف |
| SL1000 | LPV443.99.10002 | 4‑channel hannel PPA Parallel Analyzer |
| SL250 | LPV443.99.11002 | Single‑channel hannel PPA Parallel Analyzer |
5.Spectrophotometers (Frequently Quoted Alongside Colorimeters, Continuous Wavelength)
| نموذج | Catalog No. | وصف |
| DR1900 | DR1900-05C | Basic package, adapter + vials + battery set |
| DR1900 | DR1900-05K | Rechargeable + USB communication package |
| DR3900 | LPV422.99.00002 | Benchtop Visible Spectrophotometer for laboratory |
| DR6000 | LPV424.99.00002 | Benchtop UV‑Vis Spectrophotometer |
- Matching Digesters (For COD / Total Nitrogen / Total Phosphorus with DR900 / DR1900)
| نموذج | Catalog No. | وصف |
| DRB200 | LTV082.99.40001 | 15‑well Digester |
| DRB900 | LTV082.99.40002 | 25‑well Digester |
Selection Cheat Sheet
- DR900: –All-in-one multi-parameter–, for wastewater plant patrol; مُثَبَّت 4 wavelengths, no spectral scan
- DR300: –Single-parameter, low-cost pocket unit–, for tap water / circulating water spot chlorine & ozone testing
- PCII: Legacy model, stock declining; new projects specify DR300 preferentially
- DR1900: –Continuous wavelength spectrophotometer–, custom full-spectrum curves, lab & field dual use
- DR3900 / DR6000: Benchtop units, dedicated laboratory deployment