Nombre completo: APG-500C Precision Resistance (Piraní) Transmisor de vacío (Pirani Gauge)
Commonly known as Chengzhen vacuum transmitter gauge, equivalent to INFICON PSG500
Marca: Chengdu Zhenghua Electronic Instrument, Trademark: 【Chengzhen】
✅ Especificaciones clave
Principio de medición: Piraní (thermal conduction resistance vacuum gauge)
Rango de medición: 5×10⁻² Pa ~ 1×10⁵ Pa (Aire / N₂)
Fuente de alimentación: 13.5~30 VCC, Power consumption ≤2W, Max inrush current <300mA
Salida analógica: Logarithmic voltage 1.9~10.0 VDC (0~0.5V indicates filament failure)
Interfaz eléctrica: RJ45 (8-core FCC68)
Vacuum Flange: Standard KF16 (mainstream version)
Max Pressure Withstand: ≤0.2 MPa (presión absoluta)
Tiempo de respuesta: 100 EM
Requisito de montaje: Instalación vertical, keep away from heat sources and vibration sources
Medios aplicables: Aire, N₂, O₂, CO and other inert gases only; Combustible, explosive and strongly corrosive gases are prohibited
📌 RJ45 Pin Definition
| Alfiler | Definición |
| 1 | Fuerza + (13.5~30 VCC) |
| 2 | Potencia 0V (Tierra) |
| 3 | Analog Signal + |
| 4 | Transmitter Identification |
| 5 | Analog Signal – |
| 6 | Contacto de relé |
| 7 | CAROLINA DEL NORTE |
| 8 | Contacto de relé |
Relay Rating: CC 30 V. / 500mamá (Only APG500CC supports multi-point threshold control; standard APG-500C has no adjustable control points)
📎 Gas Correction Factor (Reading × C)
Air/N₂/O₂/CO=1.0;H₂=0.5;He=0.8;Ar=1.7;CO₂=0.9;Xe=3.0;Kr=2.4;Ne=1.4;Water vapor=0.5
🛠 Calibration Notes
- Preheat for at least 10 min before calibration; continuous power-on is recommended to maintain thermal balance
- ATM full-scale calibration: At atmospheric pressure, set to 1×10⁵ Pa → 10.0 VCC
- HV zero calibration: Pump down to ≤1×10⁻² Pa, set to 1×10⁻² Pa → 1.5 VCC
📌 Matching Products & Alternativas
Matching indicator: PDM-5203, PDM-520C
Import equivalent: INFICON PSG500
Modelo actualizado en la misma serie.: APG-500CC (with adjustable relay threshold)
Common failure: Platinum wire contamination, platinum wire burnout (fault output 0~0.5V). Contamination damage is excluded from warranty
📋 Typical Applications
Máquinas de recubrimiento, vacuum furnaces, leak detection systems, vacuum packaging, negative pressure processes, coarse & medium vacuum monitoring for semiconductor
APG-500C (Chengzhen/Zhenghua) Lista de verificación de solución de problemas
Condición previa: Distinguish between 【gauge failure】, 【wiring/acquisition side failure】 and 【vacuum system leakage / process issue】 first; 0~0.5V output = filament failure (alambre roto / heavy contamination). This is a hard fault, and contamination damage is not covered by warranty.
Required tools: Multimeter for RJ45 pin voltage measurement, megaóhmetro, standard vacuum source, leak detector
- Falla: Fixed output 0~0.5V (fault voltage)
Síntoma: Voltage locked at 0~0.5V, reading unchanged regardless of vacuum, indicator reports fault
Causas fundamentales:
- Burned platinum thermal wire (mas frecuente; easily damaged by repeated atmospheric shock, sudden venting or overpressure)
- Heavy oil contamination / coating dust covering the thermal wire inside the gauge, causing thermal conduction failure
- Broken wire inside RJ45 plug, poor contact on pin1/pin2 power supply
- Damaged drive PCB inside gauge
Pasos para solucionar problemas:
① Measure pin1 & pin2 with multimeter to confirm stable 13.5~30VDC supply
② Test with a confirmed good dedicated RJ45 cable
③ If power supply is normal and output remains 0~0.5V after cable replacement → sensor defective
Recurso: Replace the gauge head directly. Do not disassemble and clean the thermal wire by yourself (easy to cause permanent damage and void warranty)
- Falla: Ninguna salida en absoluto, no signal on meter, floating reading on PLC
Síntoma: No voltage at signal terminals, same as no connection
Causas fundamentales:
- Missing power supply, polaridad inversa, fusible quemado
- Broken core inside RJ45 cable, bad crimp on crystal head, broken signal core pin3/pin5
- Mismatched input impedance of acquisition terminal (required ≥10kΩ)
Pasos para solucionar problemas:
① Measure voltage pin1-pin2, verificar la polaridad (pin1=24V+, pin2=GND)
② Power the gauge separately and measure analog output directly at pin3 & pin5 to rule out rear-end wiring/module faults
③ Confirm PLC/display meter supports high-impedance voltage input; do not use current input channel
Recurso: Reemplace el cable, correct power polarity, use compatible acquisition channel
- Falla: Inaccurate reading, gran desviación, wrong value at atmosphere or high vacuum
Síntoma: Not 1E5Pa at atmospheric pressure, cannot reach expected vacuum after pumping, large deviation compared with reference gauge
Causas fundamentales:
- No preheating or calibration (≥10min preheating required)
- Gas is not air/N₂ and gas correction factor not applied (large deviation for Ar, He, H₂)
- Slight contamination or long-term drift of thermal wire
- Wrong mounting orientation (non-vertical installation, condensate/dust accumulates on thermal wire)
- Leakage at KF flange, aged KF O-ring, misaligned gasket
- Gauge inlet partially blocked by oil mist or dust
Pasos para solucionar problemas:
① Atmosphere calibration: Vent to atmosphere, stabilize for 10 mín., perform atmospheric calibration
② Pump down to ≤1×10⁻²Pa for zero calibration
③ Confirm gas type and apply correction factor
④ Check mounting posture: Instalación vertical, flange downward or lateral; avoid upside-down installation which causes liquid accumulation
⑤ Isolate gauge and perform leak test for chamber leak rate
Recurso: Recalibrate, adjust installation, replace sealing ring, add front oil baffle; replace sensor if heavily contaminated
- Falla: Reading fluctuates continuously, severe drift and instability
Síntoma: Vacuum is stable but reading jumps back and forth
Causas fundamentales:
- Large ripple / unstable power supply, shared power supply with high-power motors or inverters
- Signal cable routed together with power cable, interferencia electromagnética
- Intrinsic pressure fluctuation of system (pump pulsation, frequent valve switching, airflow disturbance)
- Slight contamination or aging of thermal wire
- Mala conexión a tierra, mixed analog signal ground and power ground
Pasos para solucionar problemas:
① Use isolated DC power supply for independent gauge power supply
② Single-end grounding for cable shield, mantener alejado de los cables de alimentación
③ Remove gauge and power on in static atmosphere; if stable at atmosphere → fault from system airflow / fuga
Recurso: Add shielding, isolated power supply, optimize grounding; add damping tube on chamber
- Falla: Permanent atmospheric reading (~10V), reading does not drop while pumping
Síntoma: Pump keeps running, gauge stays at atmospheric pressure reading
Causas fundamentales:
- Severe leakage at KF flange, missing or damaged O-ring
- Isolation valve before gauge closed, gauge separated from chamber
- Gauge inlet fully blocked by solid deposit, gas cannot enter gauge cavity
- Fallo del circuito interno, output locked at full scale
Pasos para solucionar problemas:
① Close gauge isolation valve, test gauge under vacuum separately to distinguish gauge fault or system leakage
② Visual check KF O-ring and clamp compression
③ Disassemble flange and check inlet for coating material / dust blockage
Recurso: Leak detection, open valve, clean inlet; replace sensor if blocked and thermal wire contaminated
- Falla: Abnormal relay setpoint (APG-500CC relay version only)
Síntoma: Relay does not trigger / false trigger at target vacuum
Causas fundamentales:
- Wrong threshold calibration procedure (incorrect timing of setting button)
- Relay contact overloaded (max DC30V/500mA; cannot drive high-power load directly)
- Oxidized or welded contacts
- Deriva de señal, mismatch between actual vacuum and gauge output voltage
Pasos para solucionar problemas:
① Relearn threshold following manual (hold setting pin, reach target vacuum then release)
② Test contact continuity with multimeter under no-load condition
③ Add intermediate relay for load isolation
Recurso: Recalibrate; replace whole transmitter if contacts damaged
- Falla: Frequent filament burnout (high failure rate)
Trigger list (engineering prevention focus)
- Fast frequent venting of chamber, instantaneous airflow impact on thermal wire
- Continuous deposition of process oil mist and evaporation coating material on thermal wire
- Overpressure shock (>0.2 MPa absolute)
- Combustible / corrosive gas entering gauge tube (prohibited for this model)
- Upside-down installation, condensate dripping directly onto thermal wire
Contramedidas: Install front oil trap, slow venting, add isolation valve, instalación vertical, restrict process gas
✅ Quick Self-Check Sequence (on-site recommended order)
- Medir el voltaje de suministro (pin1, pin2) → rule out reverse polarity / subtensión
- Measure analog output (pin3, pin5), first check if it is 0~0.5V fault state
- Cross-test with spare cable to separate gauge fault or cable / backend acquisition fault
- Static atmosphere test, check whether reading stabilizes at atmospheric pressure
- Isolated vacuum test to distinguish 【gauge drift】 or 【chamber leakage / pump issue】
- Calibración, check mounting posture, oil baffle and sealing
⚠️ Important Disclaimer
Thermal wire contamination, oil fouling and burnout caused by dust are not covered under warranty. The manufacturer usually only supplies replacement gauge heads instead of repair.
Do not disassemble gauge head or flush internal thermal wire with solvent, permanent damage may occur immediately.
Pirani gauge is gas-dependent and not universal vacuum gauge; correction is mandatory for Ar, He, H₂.
Chengzhen (Chengdu Zhenghua ZHAC) Lista completa de modelos
Marca: Chengzhen / Chengdu Zhenghua Electronic Instrument; grouped by category; APG Pirani transmitter (your main model) at the top
- Piraní / Resistance Vacuum Transmitters (Integrated transmitter gauge, PLC direct acquisition, equivalent to INFICON PSG500)
APG-500C: Standard Pirani resistance transmitter, RJ45, logarithmic output 1.9~10V, KF16
APG-500CC: Version with adjustable relay threshold output
FPG550: Resistencia + piezoresistive composite transmitter (amplia gama, atmosphere down to rough vacuum)
ATG-530: 1Torr range precision transmitter
- Vacuum Gauge Tubes (Separate type, requires matching indicator)
Resistance Gauge (Piraní)
ZJ-52T: Classic resistance vacuum gauge tube (glass gauge, KF, CF flange options)
Thermocouple Gauge
ZJ-51
ZJ-53B
ZJ-54D (Short-filament thermocouple, widely used in coating industry)
Piezoresistive Gauge
ZJ-1P: Piezoresistive vacuum gauge, 0.1~100kPa, suitable for near-atmosphere rough vacuum
Capacitance Diaphragm Gauge (Gas-independent, alta precisión)
ZJ-1C Series (Range options: 10t, 100t, 1000t, 0~1000Pa; 4-20mamá / 0-10salida V)
Hot Cathode Ion Gauge (B-A gauge, high vacuum)
ZJ-2
ZJ-10
ZJ-12 (B-A ultra-high vacuum gauge)
ZJ-27 (Wide-range ion gauge, mainstream for coating machines)
Cold Cathode Ion Gauge (Penning / inverted magnetron cold gauge, high vacuum, no filament)
ZJ-14 (Inverted magnetron cold gauge)
ZJ-14C (Penning cold gauge)
- Single-type Vacuum Indicators (Panel mount for control cabinet)
ZDR Resistance (Piraní) Vacuum Meter
ZDR-I-Pro: Single-channel resistance vacuum meter
ZDR-II-Pro: Dual-channel resistance vacuum meter
ZDR-X-LED: Rack-mount multi-channel resistance vacuum mete
ZDO Thermocouple Vacuum Meter
ZDO-I / ZDO-II: Soltero / dual channel thermocouple vacuum meter
ZDK Hot Cathode Ion Vacuum Meter
ZDK-I-LED, ZDK-X-LED: Soltero / multi-channel ion meter, compatible with ZJ-27/ZJ-
ZDV Cold Cathode Ion Vacuum Meter
ZDV-I-LED, ZDV-IC-LED: Dedicated indicator for cold gauge
PDM Miniature Resistance Vacuum Meter (Compacto, portable all-in-one)
PDM-5203: Panel miniature resistance vacuum meter (match APG, ZJ-52T)
PDM-520B: Handheld miniature vacuum meter (5V fuente de alimentación, field portable)
- ZDF Composite Vacuum Meter (Industria principal, resistencia + ion / par termoeléctrico / cold gauge combination, standard for coating machines)
ZDF-I-LED / ZDF-I-PRO: Par termoeléctrico + ion composite
ZDF-III-LED / ZDF-III-PRO: Resistencia (ZJ52T) + hot cathode ion (ZJ27), más común
ZDF-IV-LED / ZDF-IV-PRO: Resistencia + B-A ultra-high ion gauge (ZJ12)
ZDF-V-LED / ZDF-V-PRO: Resistencia + cold cathode cold gauge (ZJ14)
ZDF-X-LED / ZDF-X-PRO: Dual resistance + single ion
ZDF-IB / ZDF-VB / ZDF-VC: Dual thermocouple, dual resistance cold gauge, dual ion and other special composite versions
- Automatic Pressure Controller (Vacuum closed-loop pressure stabilization control)
ZDC-I: Resistencia + ion composite pressure controller
ZDRC-I: Resistance pressure controller
ZDMC-I: Capacitance gauge pressure controller
ZDFC-III: Composite vacuum pressure controller
- Accesorios a juego
ZDT-5: Electromagnetic fine-tuning vacuum valve (for pressure controller)
Ceramic sealed vacuum feedthrough electrode (vacuum electrode, vacuum feed wire)
KF/CF flange adapter, gauge cable, dedicated APG RJ45 cable
Quick Selection Reference for BOM
Rough vacuum, Adquisición de PLC, replace PSG500 → APG-500C / APG-500CC
Indicator display + rough vacuum → ZDR series + ZJ-52T
Standard coating machine composite vacuum (1e5~1e-5Pa) → ZDF-III-PRO + ZJ52T+ZJ27
Ultra-high vacuum → ZDF-IV + ZJ12
Filament-free, long-term high vacuum → ZDF-V + ZJ14C
Gas-independent high-accuracy absolute pressure → ZJ-1C capacitance gauge