Marque: Pentronique (Suède), a Nordic manufacturer of high-precision capteurs de température. Le 5810080 La série est un détecteur de température à résistance industriel Pt100 assemblé (RDT). It is often mistakenly referred to as a thermocouple in the industry; this distinction is critical.
- Explication du code modèle
5810080: Série de produits, assembled industrial platinum RTD with terminal head and protection thermowell
-001: Version / construction suffix, representing single Pt100 element, Class A accuracy, base version with standard thermowell material, probe length and thread specification
Other variants within the series include 5810080-002 et 5810080-003, commonly dual Pt100 elements, alternative probe lengths or pressure ratings.
- Spécifications techniques de base
| Article | Spécification |
| Élément de détection | Pt100 Platinum RTD (compliant with IEC60751) |
| Classe de précision | Classe A |
| Plage de température | -200℃ ~ +500℃ (depends on thermowell material) |
| Configuration du câblage | Standard 4-wire (3-wire optional). The 4-wire design eliminates lead resistance error and is preferred for high-precision measurement |
| Thermowell Material | 316L acier inoxydable (standard); Hastelloy or Titanium optional for highly corrosive media |
| Probe Construction | Solid mineral-insulated / drilled protection tube, vibration-resistant structure (Pentronic thermally conductive filling for faster response) |
| Terminal Head | Aluminium waterproof head, IP65; built-in temperature transmitter optional (4-20sortie mA) |
| Montage | Threaded installation (G1/2, M20×1.5 commonly used), compatible with protection thermowell |
| Temps de réponse | t0.5: environ. 2~8s (in liquid media, subject to probe diameter) |
| SIL | SIL2 loop assessment available (dual-element version 5810080-002) |
- Fonctionnalités principales
- Excellent stability: Class A platinum wire with annual drift as low as 0.05℃, suitable for long-term continuous process measurement
- Vibration-resistant construction: Internal metal filling reduces air gaps, withstands vibration for temperature measurement on compressor, pump and turbine casings
- Replaceable insert: Sensing element is a removable insert. Sensor core can be replaced without removing the thermowell, cutting downtime maintenance cost
- Corrosion-resistant options: Custom Hastelloy and Monel thermowells available for corrosive chemical service
- Accessoires en option: Built-in HART temperature transmitter providing direct 4–20mA signal to DCS/PLC
- Scénarios d'application typiques

Pentronique 5810080-001 Capteur de température
Thermal power, huile & gaz, chimique, paper and pharmaceutical industries; temperature measurement of pipelines, equipment casings and tank media; high-precision temperature acquisition for TSI and process control loops.
Frequently paired with Bently Nevada TSI systems for unit temperature monitoring.
- Installation & Directives de mise en service
- Installation: Protection thermowell is recommended for online replacement. Probe insertion depth ≥ 8~10 times thermowell inner diameter to ensure accurate temperature reading
- Câblage: 4-wire Pt100 with two excitation wires and two measurement wires. Must not be used with thermocouple extension wires
- Blindage: Cable shield single-point grounded inside terminal head to avoid common-mode interference; route away from power cables
- Étalonnage: Étalonnez les valeurs de résistance par rapport à la norme IEC60751 à l'aide d'un calibrateur à bloc sec ou d'un bain d'huile à température constante
- Versions antidéflagrantes: A partir de maintenant / Ex d disponible; les certificats doivent être vérifiés pour la sélection des zones dangereuses
- Liste de contrôle de dépannage courante
- Fluctuations de température: Connexions de câbles desserrées, mise à la terre multipoint du blindage, desserrage des bornes causé par les vibrations du champ
- Toujours élevé / lectures faibles: La sonde n'est pas complètement immergée dans le milieu, puits thermométrique encrassé, résistance excessive du plomb (passer en configuration 4 fils)
- Alarme de circuit ouvert: Fil de platine cassé ou dommages causés par la flexion à la racine (fréquemment observé dans des conditions de vibration)
- Dérive accrue: Vieillissement de l'élément en platine après un fonctionnement prolongé en surchauffe
- Alternative à la série & Référence de sélection
| Modèle | Construction | Application |
| 5810080-001 | Pt100 simple, Classe A | Mesure de température en un seul point pour les processus généraux |
| 5810080-002 | Double Pt100, Classe A/A | Mesure redondante, Boucle SIL2 |
| 5810080-003 | Élément unique, sonde étendue / doigt de gant haute pression | Insertion profonde, high-pressure pipelines |