Model positioning: Basic digital megohmmeter with 250-2500 V test voltage; no data storage or USB communication. The higher-end model 1537 features data storage, USB communication, fine-tunable test voltage, ramp test, and AC/DC voltage measurement.
- 핵심 전기 사양
| 목 | 흡충 1535 |
| Test Voltage Ranges | 250 다섯 / 500 다섯 / 1000 다섯 / 2500 다섯 (fixed ranges, no 100‑V‑step fine adjustment) |
| 절연 저항 범위 | 200 kΩ ~ 500 GΩ |
| Short‑circuit Current | >2 mA (>5 mA for 1537) |
| Automatic Calculation | Polarization Index (PI), Dielectric Absorption Ratio (DAR); No Dielectric Discharge (DD), no Ramp test |
| Leakage Current Measurement | 1 nA‑2 mA; Capacitance Measurement: 0.01‑2 μF |
| 안전등급 | CAT IV 600 V, IEC 61557‑5, 오염도 2 |
| Live‑circuit Alarm | High‑voltage alarm triggered above 30 V; auto‑discharge and residual‑voltage display upon test completion |
| 전원공급장치 | 8 × AA (LR6) alkaline batteries; ≥1300 tests at 2500 V; ≥6500 tests at 250 V |
| 치수 / 무게 | 211 × 93 × 184 mm, 1.3 kg |
| 온도 범위 | 운영 중: ‑10℃ ~ +50℃; 저장: ‑20℃ ~ +60℃; Non‑condensing; Altitude ≤2000 m |
| 유입 방지 | IP40; Backlit LCD; No internal storage, No USB port |
- 모델 비교: 1535 대 1537
| 기능 | 1535 | 1537 |
| Fixed voltage ranges: 250/500/1000/2500 다섯 | ✅ | ✅ |
| 250‑2500 V fine adjustment in 100 V steps | ❌ | ✅ |
| Automatic PI / DAR calculation | ✅ | ✅ |
| Dielectric Discharge (DD), Ramp voltage test | ❌ | ✅ |
| ACV / DCV and general resistance measurement | ❌ | ✅ |
| 데이터 저장, tagging, USB upload to PC | ❌ | ✅ |
| Short‑circuit current | >2 mA | >5 mA |
- Standard Packaging List
Standard Deliverables:
- 흡충 1535 Main Unit
- 3-colour test leads (빨간색, 검은색, 녹색) with alligator clips
- Red-black test probes
- Soft carrying case
- 사용 설명서 & quick-start guide
❗ Not included: USB cable, PC 소프트웨어, hard-shell tool case (order separately); 8-AA batteries to be provided by user.
- 일반적인 응용 분야
- Insulation testing for motor & generator windings (low-voltage and medium-voltage motor pre-testing)
- Insulation preventive testing for transformers, cables and busbars
- Insulation test for DC side of PV arrays (2500 V range suitable for photovoltaic strings)
- On-site electrical maintenance, fault troubleshooting, pre-/post-overhaul insulation evaluation
- Plant equipment spot-check; handheld on-site use for scenarios without report-generation requirements
👉 Upgrade to Fluke 1537 if you need data logging, report export, ramp withstand test or fine voltage adjustment.
- Selection Pitfalls & Hard-rule Warnings
- 1535 has no storage and cannot export reports via PC. PI / DAR values must be manually recorded on-site; do not expect CSV export after purchase.
- Short-circuit current is only >2 mA. For large-capacity long-length cables, charging time is longer and readings stabilize slower compared with the 1537 (5 엄마).
- Never perform insulation tests on live circuits. The instrument triggers alarm above 30 V, but full power-off and adequate discharge of the unit-under-test are mandatory.
- 결로금지. Measurement distortion occurs under high-humidity conditions with dew formation.
- Derate operation above 2000 m altitude; full-range 2500 V testing is not recommended.
- Always wait for automatic discharge completion and verify zero residual voltage before disconnecting test leads.
- Compatible Alternatives
동일 브랜드 업그레이드: 흡충 1537 (identical form factor, adds storage, USB and ramp test)
Equivalent competitors: Megger MIT2500; HIOKI IR3455-30
전체 비교: 흡충 1535 vs Fluke 1537 절연저항 시험기
These two models share identical enclosure, 치수, weight and base measuring range. Key differences lie in short-circuit current, voltage tuning, advanced test functions, data-logging communication and additional multimeter features.
| 비교항목 | 흡충 1535 | 흡충 1537 |
| Fixed test voltage ranges | 250 다섯 / 500 다섯 / 1000 다섯 / 2500 다섯 | 250 다섯 / 500 다섯 / 1000 다섯 / 2500 다섯 |
| Fine‑step adjustable voltage | ❌ Only four fixed ranges, no fine‑tuning | ✅ 250‑2500 V adjustable in 100 V increments |
| Insulation resistance measuring range | 200 kΩ‑500 GΩ | 200 kΩ‑500 GΩ |
| Short‑circuit current | >2 mA | >5 mA (faster charging & stabilized readings for large‑capacity cables) |
| Polarization Index (PI) / Dielectric Absorption Ratio (DAR) | ✅ Auto‑calculated | ✅ Auto‑calculated |
| Dielectric Discharge (DD) | ❌ Not supported | ✅ Supported |
| Ramp withstand‑voltage test | ❌ Not supported | ✅ 100 V/s linear ramp with breakdown‑voltage indication |
| ACV / DCV and general resistance measurement | ❌ No multimeter function | ✅ Measures AC/DC voltage & general resistance |
| 데이터 저장 & tagging | ❌ Real‑time reading only | ✅ Storage, tagging & scheduled‑test recording |
| USB interface & PC report‑export software | ❌ No USB port, no export capability | ✅ Mini‑USB, report generation with bundled PC software |
| Leakage current / Capacitance measurement | ✅ 1 nA‑2 mA; 0.01‑2 μF | ✅ 1 nA‑2 mA; 0.01‑2 μF |
| Safety rating | CAT IV 600 V | CAT IV 600 V |
| 힘 / 무게 / 기계 설계 | 8 × AA batteries; 1.3 kg; IP40 | 8 × AA batteries; 1.3 kg; IP40 |
Key Difference Explanations
- Short-circuit current (easily overlooked parameter)
그만큼 1537 delivers >5 mA short-circuit current: for long cables and large-capacitance motor windings, charging completes rapidly without long waiting for stable readings.
그만큼 1535 provides only >2 mA. Long stabilization time is required for high-capacitance test objects; it suits small-to-medium motors and short cables.
- Exclusive Ramp test for 1537
Performs withstand-voltage scanning with gradual voltage rise to pinpoint insulation weak points. Widely used in motor overhaul, high-voltage pre-test and incoming-product inspection. 그만큼 1535 outputs fixed high voltage only and cannot execute gradual-ramp tests.
- Data logging & report output
흡충 1535: No onboard storage; all readings must be manually transcribed. Suitable for spot-checks, fault-finding and work without formal reporting requirements.
흡충 1537: Stores multiple test sets, exports data via PC and generates reports. Ideal for third-party inspection, power-industry preventive testing, PV site acceptance and archive-oriented workflows.
- Built-in AC/DC voltage measurement
그만큼 1537 integrates basic multimeter functions to verify residual voltage on-site without carrying an extra multimeter, improving convenience in substation work.
선택 지침
✅ Choose Fluke 1535:
Routine plant spot-check and fault diagnosis
Small-to-medium motors and short-length cables
No data-storage or formal-report requirement, manual paper-based logging
Budget-constrained projects
✅ Choose Fluke 1537:
Formal test-report generation and data archiving
Test objects include long-distance power cables, large-capacity motors and large transformers
Advanced insulation diagnostics: Ramp test & Dielectric Discharge for aged-insulation assessment
Non-standard test-voltage requirements (예를 들어. 1700 다섯, 2100 다섯)
All-in-one instrument for insulation test plus on-site voltage measurement
Packaging Differences
흡충 1537: Main unit, test leads & probes, soft case; USB cable and PC software CD included. Hard-shell tool case sold separately.
Both models require user-supplied 8-AA batteries.
Hard-rule Notes
- Both instruments require fully power-off test objects. The 30-V live-circuit alarm serves as a reminder and cannot replace formal power-cut & voltage-verification procedures.
- Avoid applying the 1535 to ultra-capacitive long cables: long charging cycles and reading drift may occur.
- 그만큼 1537 stores data locally without onboard printing; reports must be generated after importing datasets to PC software.
흡충 1535 vs 1537|Selection Conclusion for Power-industry Applications
For power-sector applications (preventive testing, substations, distribution networks, third-party accredited testing agencies), select the Fluke 1537 기본적으로. 그만큼 1535 is only acceptable for simple fault-finding without formal data archiving.
Power-field Comparison (aligned with DL/T preventive-test specifications)
| Power‑sector Requirement | 흡충 1535 | 흡충 1537 | Practical Impact for Field Work |
| Short‑circuit current | >2 mA | >5 엄마 | 긴 케이블용, large transformers and generator windings (high‑capacitance DUTs), 1537 achieves fast charging and stable readings and cuts waiting time. 1535 exhibits slow charging and reading drift on high‑capacitance equipment. |
| 데이터 저장 + USB export | ❌ No storage, only on‑site reading | ✅ Onboard storage, PC export | Preventive testing and condition‑based maintenance demand traceable raw datasets, formal reports and equipment‑insulation archives. 그만큼 1535 only supports manual paper logging without traceability. |
| 100‑V‑step adjustable voltage | ❌ Four fixed ranges only | ✅ 250‑2500 V step tuning | Enables non‑standard test voltages for diagnosis of legacy / aged equipment. |
| Dielectric Discharge (DD) & Ramp test | ❌ Not supported | ✅ Supported | Ramp scanning detects latent insulation defects and prevents equipment damage from direct full‑voltage stress. DD quantifies ageing of old insulation systems. |
| Native AC/DC voltage measurement | ❌ Absent | ✅ Available | Residual‑voltage verification directly on the tester; eliminates the need for an extra multimeter at substations. |
| PI / DAR automatic calculation | ✅ Supported | ✅ Supported | Both comply with PI/DAR test requirements for transformers, motors and cables specified in DL/T474.1‑2018. |
Scenario-based Recommendations
✅ Mandatory selection: 흡충 1537 (Most power-industry scenarios)
- Power-utility substation maintenance, distribution-network preventive testing and equipment condition evaluation
- Accredited third-party power-testing services requiring formal traceable test reports
- Test subjects include long-distance power cables, high-capacity distribution transformers and generators / 대형 모터
- Advanced insulation diagnostics: ramp-voltage scanning and Dielectric Discharge assessment for aged insulation
- Digital-operation workflows with PC-imported datasets linked to equipment asset ledgers
For high-capacitance test items in power preventive tests, 5 mA short-circuit current is a critical performance index. The 2 mA output of the 1537 results in prolonged unstable readings for long-length cables and reduces field efficiency.
✅ Acceptable selection: 흡충 1535 (Limited scenarios only)
- Distribution-network emergency repair and on-site fault troubleshooting without formal preventive-test reporting or long-term archiving
- Test objects are small-capacity equipment and short cables with low capacitance
- 예산 제약, qualitative insulation judgement only; all PI/DAR values manually logged on paper.
Critical Selection Warnings
- The Fluke 1535 cannot export data. Not recommended for accredited third-party testing or acceptance projects where report traceability is mandatory.
- Maximum test voltage for both instruments is 2500 V. Neither can perform 5000-V insulation tests for ≥10 kV main transformers. Select Fluke 1555 / 1550C for 5-kV-range requirements.
- Both units feature CAT IV 600 V safety rating, auto-discharge and residual-voltage indication with identical safety performance.
조달 조언
Standard power-team configuration: Primary tester = Fluke 1537. Supplement with a small quantity of 1535 units for emergency-repair crews for fast fault diagnosis.
Hard-shell carrying cases and extended-length test-lead kits are not standard accessories for the 1537. These are strongly recommended for power-site deployment.
접촉기,회로 차단기,태양광 인버터,전기 계량기,태양 전지