HACH DR900 휴대용 색도계

Catalog No.: 9385100 (CN Chinese Version)– 원칙: LED colorimetric method, used for rapid on-site water quality testing with Hach preformed reagents ✅ Core Specifications Light source: 주도의, 4 fixed wavelengths: –420 / 520 / 560 / 610 nm–, wavelength accuracy ±1 nm Detector: Silicon photodiode Photometric range: 0~2 Abs; photometric accuracy ±0.03 Abs Readout modes: ...

  • 제품 세부정보

Catalog No.: 9385100 (CN Chinese Version)–

원칙: LED colorimetric method, used for rapid on-site water quality testing with Hach preformed reagents

✅ 핵심사양

광원: 주도의, 4 fixed wavelengths: –420 / 520 / 560 / 610 nm–, wavelength accuracy ±1 nm

탐지기: Silicon photodiode

Photometric range: 0~2 Abs; photometric accuracy ±0.03 Abs

Readout modes: Concentration, Absorbance (Abs), % Transmittance

Pre-programmed methods: –90 built-in methods–; 최대 지원 10 custom user curves

데이터 저장: 500 GLP-compliant records (시간, sample ID, user ID), export via Mini USB

Sample cell: 1-인치 (25 mm) round sample vial; 호환 가능 16 mm digestion tubes (COD tubes) via adapter

진입 보호: –IP67 (when the cell lid is closed)–, shock-resistant and drop-proof

전원공급장치: 4 × AA 배터리; backlit graphic display (240×160), supports Chinese plus 21 other languages

치수: 231 × 96 × 48 mm; 대략적인 무게. 0.6 kg

작동 온도: 10~40 ℃, humidity ≤80% non-condensing

✅ Typical Measurable Parameters

Free chlorine, total chlorine, ammonia nitrogen, COD, total phosphorus, total nitrogen, 6가 크롬, total chromium, 구리, 철, manganese, zinc, fluoride, 흐림, 알류미늄, barium, 등.

적용 가능한 시나리오: Municipal wastewater, 수도물, circulating cooling water, 산업폐수, surface water, aquaculture, 등.

✅ 표준 액세서리

DR900 main unit, 2 × 25 mm glass sample vials, 2 × 1 cm plastic cuvettes, 16 mm COD tube adapter, 4 × AA 배터리, Mini USB cable, multilingual manual, CD documentation

✅ 주요 기능

  1. Portable field unit, IP67 dust and water resistance, ideal for field work and wastewater plant inspections
  2. Favorites function for one-touch access to frequently used methods
  3. Supports blank correction and standard correction to offset reagent batch variations
  4. Data export to PC for environmental audit and record-keeping compliance
  5. Designed exclusively for Hach pre-packaged powder pillows and Test’N Tube digestion reagents; not suitable for arbitrary use with self-prepared reagents

✅ Comparison Reference

DR900: –4 fixed-wavelength colorimeter–, cost-effective for rapid field testing; no spectral scanning capability

DR1900: Portable spectrophotometer with continuously adjustable wavelength, suitable for both laboratory and field use

✅ Common Spare Parts & 소모품

Sample vials, cuvettes, 16 mm adapter, 배터리, dust cover, matched preformed reagents (COD, ammonia nitrogen, total phosphorus, 등.)

HACH DR900|Common Parameters – Method Numbers – Range List (High-frequency for Municipal Wastewater / Tap Water / 순환수)

메모: –Different ranges for the same parameter correspond to different method numbers; the matching preformed reagent must be used.Unit: mg/L; chlorine expressed as Cl₂; ★ denotes EPA-approved methods for environmental reporting

매개변수Method No.범위Reagent Format비고
COD (Low Range LR)80000.7~40TNT TubeSurface water, low‑concentration wastewater
COD (Mid Range MR)80003~150TNT Tube★ Most common for municipal wastewater
COD (High Range HR)800020~1500TNT TubeIndustrial wastewater
COD (Ultra‑high Range)8000200~15000TNT TubeHigh‑strength wastewater
Ammonia Nitrogen (Nessler LR)80380.02~2.50Powder PillowTap water, surface water
Ammonia Nitrogen (Nessler HR)80380.1~10.0Powder PillowWastewater plant effluent
Ammonia Nitrogen (Salicylate Method)100310.01~0.50Powder PillowLow concentration, less interference
Total Phosphorus (Ammonium Molybdate LR)81900.02~2.50Powder Pillow폐수
Total Phosphorus (TNT)102090.06~3.50TNT TubeDigestion required
Orthophosphate80480.02~2.50Powder PillowNo digestion required
Total Nitrogen100711~16TNT TubePersulfate digestion, ★
Free Chlorine (DPD)80210.02~2.00Powder PillowTap water inspection
Free Chlorine (DPD HR)100690.1~10.0Powder PillowHigh‑chlorine disinfection water
Total Chlorine (DPD)81670.02~2.00Powder PillowCirculating water
Total Chlorine (DPD HR)100700.1~10.0Powder PillowHigh chlorine
Chlorine Dioxide101260.04~5.00Powder PillowChlorine dioxide disinfection
Hexavalent Chromium Cr⁶⁺80230.01~0.60Powder PillowIndustrial wastewater
Total Chromium80240.01~0.60Powder PillowDigestion required
Total Iron (FerroVer)80080.02~3.00Powder PillowCirculating water, boiler water
Ferrous Iron81470.02~3.00Powder PillowNo digestion required
Manganese81490.01~2.00Powder PillowGroundwater
구리85060.04~5.00Powder PillowIndustrial water
Zinc80090.05~3.00Powder PillowElectroplating wastewater
Fluoride80290.1~2.0SPADNS ReagentSurface water
Total Hardness (as CaCO₃)80302~1000Powder PillowCirculating water
Turbidity86600~1000 NTU직접 측정No reagent required, sample vial only
알류미늄80120.008~0.80Powder PillowAluminum residual in tap water
Barium102515~100Powder PillowOilfield produced water
Bromine80160.05~4.50Powder PillowBromine disinfection
Phenol80440.002~0.12Powder PillowVolatile phenol
TOC101290.5~25TNT TubeTotal Organic Carbon

Quick Operating Notes

  1. For one method number (예를 들어. 8000 for COD), –range differentiation is identified by reagent tube barcode / 범위–. Select the same method number on the instrument and change reagents for different ranges.
  2. TNT digestion tubes must be used with the 16 mm adapter in DR900; powder pillow tests generally use the 25 mm round sample vial.
  3. Turbidity test 8660: Fill vial directly with sample, use pure water for blank; no reagent needed.
  4. Digestion-required parameters (total nitrogen, TNT total phosphorus, total chromium): DRB200 / DRB900 digester required.
  5. 90 methods are preloaded in total; the table above covers more than 30 commonly used engineering methods. Niche methods for boron, hydrazine, sulfide, 등. can be extracted separately.

HACH DR900 Troubleshooting Checklist

Preface: The DR900 is a fixed 4-wavelength LED colorimeter. Most faults originate from –optical path, sample vial, adapter, 전원 공급 장치, data and operation–. Diagnose by error code / symptom and troubleshoot from simple to complex.

  1. Power-on Faults

-1. 전원이 켜지지 않습니다 / Black screen on startup

배터리 부족: Replace with fresh AA alkaline batteries; carbon batteries are not recommended due to unstable voltage

Poor battery contact: Remove batteries, clean metal contacts in battery compartment, check for bent terminals

Battery polarity reversed: Match +/marking inside compartment

Corrosion from water ingress in battery compartment: Wipe stains and green corrosion with anhydrous alcohol and fully dry before retest; heavy corrosion requires factory repair

Power note: USB connection is for data export –오직; cannot power the unit for startup–

-2. Restarts repeatedly / Shuts down automatically after power-on

Aged high-internal-resistance batteries: 부하시 전압 강하; retest with new batteries

Ambient temperature out of specification: <10 ℃ 또는 >40 ℃; battery capacity drops sharply at low temperature

Loose battery contacts: Gently pry contact tabs outward to improve contact pressure

  1. Optical Path & Reading Abnormalities (가장 빈번한)

-3. Unstable readings, continuous fluctuation, heavy drift

Fingerprints, water stains or reagent residue on outer vial surface: Wipe outer wall with lint-free soft cloth; only grip the frosted zone

Scratched or worn sample vial: Scattering from scratches; replace vial directly

Air bubbles in sample: Allow to settle; avoid vigorous shaking. Tap TNT COD tubes gently to release bubbles

Turbid sample or sediment: Let sample settle or wait for full color development per reagent instructions

Misaligned vial seating: Align vial with slot marker and notch with positioning feature

Light source aging: Abnormally high blank absorbance, poor repeatability on identical sample; suspect LED degradation

Direct strong ambient light: Shield sample cell from sunlight, which interferes with optical measurement

-4. Blank calibration failure / Blank Fail error

Impure blank water: Use ammonia-free / COD-free pure water; tap water or mineral water cannot be used as blank

Contaminated blank vial: Residual carryover from high-concentration prior tests; soak in acid or replace

Mismatched paired vials for blank and sample: Hach factory vials are recommended to be used as matched pairs

Spilled liquid inside cell / dirty optical window: Open sample compartment, clean quartz window with lint-free cloth dampened with pure water and dry fully

Wrong method selected: Mismatched method and reagent will cause blank failure

-5. Negative readings

Blank vial and sample vial swapped

Sample concentration below method detection limit

Dirty / scratched vial making blank more opaque than sample

Expired / degraded reagent with no color formation

Drifted calibration curve: Perform blank recalibration

-6. Over-range reading (OL / Over Range)

Sample concentration exceeds upper limit of selected method: Dilute sample manually, select higher-range method and reagent

Wrong reagent dosage or reversed addition sequence

Incomplete digestion for COD / total nitrogen: Insufficient digestion temperature or hold time

메모: DR900 has no automatic dilution function; manual dilution and retest required

-7. Optical error / Lamp Error

Sample lid not fully closed, poor magnetic switch contact

Optical window covered with liquid or crystals → clean window

Failed LED light source: If error persists after cleaning, battery replacement and lid closure, return for LED assembly replacement

Water ingress in compartment and damp optical PCB: Discontinue use, dry for 24 hours and retest

  1. TNT Digestion Tube & 16 mm Adapter Related Errors

-8. Inaccurate or erratic COD / total nitrogen results with TNT tubes

Adapter not fully seated or tilted, causing inclined tube

Condensation droplets or reagent crystals on outer wall after digestion: Wipe completely

Tube label / graduation mark blocking optical path: Label must be kept away from light path

Sediment settled across light path: Invert gently and let sediment settle below optical path

Debris or broken glass inside adapter cavity: Clean adapter interior

Never insert a 16 mm digestion tube directly into the 25 mm cell without the adapter

  1. 작업, 방법 & Reagent Issues

-9. Large deviation between DR900 results and laboratory reference

  1. Verify three items: –method number, reagent batch, reaction hold time(DPD chlorine fades over time; Nessler ammonia nitrogen drifts with extended standing)
  2. Check interferences: Sulfide, heavy metals, oxidants and high salinity cause interference (refer to Hach reagent method interference sheets)
  3. Digestion tests: Confirm DRB200 / DRB900 digester temperature and timing accuracy
  4. Expired or moisture-damaged reagents: Caked powder pillows, discolored TNT tubes must be discarded
  5. Sample holding time exceeded: Unpreserved analytes degrade rapidly (especially chlorine and ammonia nitrogen)

-10. Required test method missing

펌웨어 버전 확인; older firmware contains fewer built-in methods. Contact Hach for firmware upgrade support

The DR900 has a fixed library of 90 embedded methods. –Spectral scan methods cannot be added freely–, only up to 10 custom user curves are supported

Lost custom curves: User curves are stored locally; factory reset deletes all custom curves

  1. 데이터 & USB Communication Faults

-11. USB connection failure, 데이터를 내보낼 수 없습니다

Must use Mini USB data cable; many charging cables lack data conductors. Use original Hach cable

Missing PC driver: Hach uses standard HID class, no special driver required. Better compatibility with Win10/11; some stripped system builds fail recognition

Instrument must be powered on when connected via USB

500-record storage full: Export records then delete old entries. GLP records support single deletion only, no bulk erase

Export software: Hach DataCom; generic serial terminal software is not compatible

-12. Lost records / Incorrect clock time

Depleted internal clock battery (onboard coin cell): Time resets after power-off; return for clock battery replacement and time setup

Factory reset executed accidentally, clearing custom curves and stored records

  1. 하드웨어, Ingress & Water Damage

-13. Liquid spillage / water entry in sample compartment

IP67 rating applies –only when lid is fully closed–. No water protection when lid is open

Immediately remove batteries after water ingress, leave lid open to air dry ≥24 hours in cool shaded area. Do NOT apply hot air (damages LED and PCB)

If short circuit or corrosion signs are present, avoid repeated power cycles and send for service

-14. Abnormal screen, garbled display, missing segments

Slow LCD response at low temperature: Normal behaviour; restore after warming

Physical impact and casing deformation: Loose ribbon cable or damaged display; return for repair

  1. 메뉴 & 구경 측정

-15. Instrument prompts for standard calibration

Routine DR900 operation only requires –blank measurement per test–. Standard calibration is only needed for custom user curves

Factory embedded methods normally do not require standard recalibration. If persistent bias occurs, verify with Hach certified check standard

  1. On-site Rapid Troubleshooting Sequence (Field Emergency)
  2. Install new batteries → 2. Clean sample vials and optical window → 3. Re-run blank calibration → 4. Check sample for bubbles and sediment → 5. Confirm method number, reagent and reaction time → 6. Inspect lid magnet and adapter seating → 7. Check temperature and strong light exposure → 8. Verify with standard sample to isolate instrument vs sample/reagent error
  3. 금지된 작업

Do not clean optical windows with acetone or aggressive solvents (coating damage)

Do not scrub vials with abrasive paper or hard material (permanent scratches)

Do not insert hot / pressurized TNT tubes; cool fully to ambient temperature before measurement

Do not disassemble the housing; disassembly voids warranty

HACH Colorimeter / Portable Spectrophotometer Model & Catalog List (Mainstream China SKUs)

구별: –Colorimeter (fixed LED wavelength)대 –Spectrophotometer (continuously tunable wavelength)–; PCII is legacy, DR300 is PCII replacement; DR900 multi-parameter fixed wavelength; DR1900 and above are spectrophotometers often grouped together for quotation

메모: CN = Chinese version; catalog numbers are standard Hach China SKUs; OEM part numbers available for foreign trade

  1. DR900 Multi-parameter Portable Colorimeter (4 fixed wavelengths)
모델카탈로그 번호.설명
DR900 (Standard Chinese Version)9385100본체 + 25 mm vials, 1 cm cuvettes, 16 mm COD adapter, 배터리, USB 케이블, 수동, 90 built‑in methods, IP67
  1. DR300 Pocket Single-parameter Colorimeter (PCII successor, IP67, Bluetooth optional)
모델카탈로그 번호.Measured Parameter
DR300LPV445.80.00110Free & Total Chlorine (DPD)
DR300LPV445.97.01110Bromine Br₂
DR300LPV445.97.02110Nitrate
DR300LPV445.97.03110Dissolved Oxygen
DR300LPV445.97.04110Ozone
DR300LPV445.97.06110Orthophosphate
DR300LPV445.97.09110Zinc
DR300LPV445.97.10110Molybdenum
DR300LPV445.97.25110알류미늄
DR300LPV445.97.26110Monochloramine / Free Ammonia
DR300LPV445.97.40110Ammonia Nitrogen
DR300LPV445.80.51110Chlorine Dioxide
DR300LPV445.97.62110Free / Total Chlorine Mid‑High Range
DR300LPV445.97.22110Total Iron (FerroVer)
DR300LPV445.97.16110Iron TPTZ
  1. PCII (Pocket Colorimeter II, legacy single-parameter, phased out for DR300)
모델카탈로그 번호.매개변수
PCII5870000Free / Total Chlorine DPD
PCII5870050Chlorine Dioxide
PCII5870040Ozone
PCII5870030Bromine
PCII5870020철
PCII5870010Manganese
  1. SL Series Portable Parallel Analyzer (PPA, multi-channel rapid field)
모델카탈로그 번호.설명
SL1000LPV443.99.100024‑channel hannel PPA Parallel Analyzer
SL250LPV443.99.11002Single‑channel hannel PPA Parallel Analyzer

5.Spectrophotometers (Frequently Quoted Alongside Colorimeters, Continuous Wavelength)

모델카탈로그 번호.설명
DR1900DR1900-05CBasic package, adapter + vials + battery set
DR1900DR1900-05KRechargeable + USB communication package
DR3900LPV422.99.00002Benchtop Visible Spectrophotometer for laboratory
DR6000LPV424.99.00002Benchtop UV‑Vis Spectrophotometer
  1. Matching Digesters (For COD / Total Nitrogen / Total Phosphorus with DR900 / DR1900)
모델카탈로그 번호.설명
DRB200LTV082.99.4000115‑well Digester
DRB900LTV082.99.4000225‑well Digester

Selection Cheat Sheet

  1. DR900: –All-in-one multi-parameter–, for wastewater plant patrol; 결정된 4 wavelengths, no spectral scan
  2. DR300: –Single-parameter, low-cost pocket unit–, for tap water / circulating water spot chlorine & ozone testing
  3. PCII: 레거시 모델, stock declining; new projects specify DR300 preferentially
  4. DR1900: –Continuous wavelength spectrophotometer–, custom full-spectrum curves, 랩 & field dual use
  5. DR3900 / DR6000: Benchtop units, dedicated laboratory deployment