Marke: LUTRON, Taiwan
LUTRON VB-8200 User Manual PDF
It is a handheld industrial vibration meter for equipment inspection, vibration severity measurement of Motoren, Pumpen und Ventilatoren. It only measures vibration velocity and acceleration, without displacement measurement (VB-8220 is the version with displacement function).
✅ Kernfunktionen
- Messung: Vibration velocity and vibration acceleration; supports RMS (Root Mean Square) and Peak reading.
- Reading functions: Data Hold, Maximal / Minimum value recording.
- Kommunikation: Isolated RS232 serial output, compatible with host computer software for data acquisition and recording.
- Large LCD display, low battery indicator and auto power-off.
- Separated piezoelectric vibration probe, supplied with magnetic base for easy attachment to equipment housing for measurement.
- Complies with ISO 2954 for vibration transducers, used for equipment vibration severity assessment per ISO 2372.
📋 Technical Specifications
| Artikel | Parameter |
| Vibration Velocity Range | 0.5 ~ 199.9 mm/s (voller Maßstab 200 mm/s) |
| Acceleration Range | 0.5 ~ 199.9 m/s² (voller Maßstab 200 m/s²) |
| Frequenzgang | 10 Hz ~ 1 kHz |
| Sampling Time | Ca. 1 zweite |
| Betriebsumgebung | 0~50℃, humidity <80%RH |
| Stromversorgung | 9V battery (006P/PP3) |
| Gewicht | 230G (excluding probe), probe: 38G |
| Zertifizierungen | ISO9001, CE, IEC1010 |
🛠️ Typische Anwendungen
Routine inspection of rotating machinery: Motoren, Kreiselpumpen, Fans, gearboxes, Luftkompressoren
Preliminary screening of faults: rotor unbalance, Fehlausrichtung, bearing looseness, foundation looseness
Factory predictive maintenance, equipment acceptance, vibration severity classification (ISO2372)
Laboratory and factory outgoing simple vibration testing
📦 Standard Accessories
Haupteinheit, separated vibration sensor, magnetic mounting base, 9V battery, hard carrying case. Host computer acquisition software is optional.
⚠️ Model Differentiation (Easily Confused Models)
VB-8200: Velocity + Acceleration, no displacement, RS232
VB-8220: Velocity + Acceleration + Peak-to-Peak displacement (2mm p-p)
VB-8212: Aktualisierte Version, some versions with temperature measurement
📌 Operation Notes
- Magnetic base is suitable for flat metal surfaces; use probe tip contact measurement for rough surfaces.
- Lower frequency limit at 10Hz, not suitable for low-frequency large-displacement equipment (z.B. large steam turbines). Low-frequency measurement requires dedicated low-frequency vibration meters.
- Only for simple inspection and severity evaluation. Spectrum analysis and bearing envelope demodulation require professional vibration analyzers (not this unit).
❗ Allgemeine Fehlerbehebung
- Fluctuating readings: Broken probe cable, poor contact of magnetic base, rusted measuring surface.
- Zero reading: Damaged probe or loose probe connector.
- Failed PC data upload: RS232-to-USB driver fault, Nicht übereinstimmende Kommunikationsparameter.
- Low battery warning: Insufficient battery voltage leads to degraded reading accuracy.
Alternative Auswahlreferenz
Inländische Alternativen: Anbo VBT series, VM series vibration meters.
Importierte Äquivalente: BK, IMV portable vibration meters.
Core Application of LUTRON VB-8200 Vibration Meter
Zusammenfassung: Handheld simple vibration measuring instrument, used to measure vibration severity of rotating machinery for routine predictive maintenance, judge whether equipment vibration exceeds threshold and screen early faults.
Measurable Physical Quantities
Vibration Velocity VEL: mm/s (am häufigsten verwendet; ISO2372 equipment rating uses this parameter)
Vibration Acceleration ACC: m/s²
Reading modes: RMS / Peak; with max/min recording and data hold
❗No support for vibration displacement, spectrum analysis, demodulation or time-domain waveform. Cannot perform precision fault diagnosis; only severity evaluation.
Main Measurement Objects (Factory Inspection)
Motoren, Kreiselpumpen, Fans, Kühlturmventilatoren, gear reducers, Luftkompressoren
Small rotating production equipment
Types of Faults for Preliminary Screening (Qualitativ)
- Rotor unbalance: Uneven mass of rotating parts causing high vibration
- Misalignment: Coupling misalignment between motor and pump
- Bearing deterioration: Bearing wear leads to vibration increase (only detects higher vibration level; cannot identify specific bearing fault type)
- Loose anchor / foundation, insufficient housing rigidity
- Imbalance caused by fouling and deposit buildup on blades or impellers
Typical Working Scenarios
- Routine equipment inspection: Inspectors measure vibration at fixed measuring points periodically, record values and compare historical data to monitor vibration rising trend.
- New equipment / overhaul acceptance: Measure vibration after equipment maintenance and judge compliance against ISO2372 standard.
- Preliminary fault screening: Rapid vibration measurement when abnormal noise occurs to assess vibration severity. If vibration exceeds limit, use professional vibration analyzer for further spectrum diagnosis.
- Datenprotokollierung: Built-in RS232 serial port connects to PC for vibration data collection and inspection report generation.
Einschränkungen (Critical for Avoiding Misuse)
- Frequency range: 10Hz ~ 1kHz, not applicable to low-frequency large units (steam turbines, large low-speed motors)
- Only severity measurement, no spectrum function to distinguish fault types. It only tells you “vibration is high”, and cannot easily tell whether it is unbalance or bearing failure.
- Not suitable for high-frequency impact measurement of precision bearings (no envelope demodulation function).
- Only measures casing surface vibration, not shaft vibration (shaft vibration requires eddy current probes).
Simple Example
For a water pump, measure vibration velocity at bearing position weekly.
ISO2372 standard: Within 10mm/s = Good; above 11.2mm/s requires monitored attention; above 18mm/s is dangerous and equipment shall be shut down for maintenance. VB-8200 reads this mm/s value.
Full Model List of LUTRON VB-82xx Series Vibration Meters
Gleiche Serie (VB82xx inspection vibration meters, 10Hz~1kHz, separated piezoelectric probe, ISO2954). VB-8200 is the base model. Models differ by measurement parameters, data storage and channel count.
| Modell | Measurable Parameters | Hauptmerkmale | Positionierung |
| VB‑8200 | Vibration velocity, Beschleunigung | Basismodell; RS232; no displacement; MAX/MIN, HOLD; no SD card | Routine equipment inspection (your selected model) |
| VB‑8201HA | Vibration velocity, Beschleunigung | Simplified VB‑8200, ähnliche Funktionen, slight housing difference, RS232 output | Economical inspection |
| VB‑8220 | Velocity, Beschleunigung, peak-to-peak displacement | Upgraded from VB‑8200 with displacement measurement; optional RS232 | Inspection when displacement reading is required |
| VB‑8210 | Velocity, Beschleunigung | Same hardware as VB‑8200, panel/firmware fine-tuning, supports imperial unit, RS232 | Export version for overseas market |
| VB‑8212 | Velocity, Beschleunigung, some versions with temperature | Early old model, limited stock in market, gradually replaced | Spare parts for legacy projects |
| VB‑8206SD | Velocity, Beschleunigung, displacement; SD card local storage | Hintergrundbeleuchtetes LCD, auto data logging with adjustable sampling interval, RS232 | Long-term on-site data acquisition without continuous PC connection |
| VB‑8230SD | Velocity, Beschleunigung, displacement; Dual channel | Two probes simultaneous acquisition, SD card logging, dual-channel comparison measurement | Comparative test of bearing ends / motor-pump alignment |
| BVB‑8207SD | Velocity, Beschleunigung, displacement; Four channel | 4-channel vibration probe simultaneous measurement; SD card + RS232/USB; larger main unit | Multi-point synchronous acquisition for unit inspection |
| PVB‑820 | Velocity, Beschleunigung | Pen-type integrated vibration meter, built-in probe without separated sensor, quick simple screening | Simple handheld pen-style inspection |
| VT‑8204 | Velocity, Beschleunigung | Basic integrated vibration meter, no separated probe, low-cost entry model | Quick simple vibration measurement |
Kompatible Sensoren (Interchangeable within Series)
VB-83: Standard piezoelectric probe for VB82xx series, equipped with magnetic base, BNC connector. Interchangeable on VB-8200 / VB-8206SD / BVB-8207SD and other models.
✅ Quick Selection Guidelines
- Displacement required? VB-8200/8201HA/8210/PVB820: no displacement; 8220, 8206SD, 8230SD, BVB-8207SD support displacement.
- SD local storage required? Models with suffix SD support SD card logging; models without SD only support real-time online acquisition via RS232.
- Channel quantity: Einkanal: 8200/8220/8206SD; Dual channel: 8230SD; Four channel: BVB-8207SD.
- Integrated vs Separated probe: PVB-820, VT-8204: integrated probe; all other VB-82xx: separated BNC probe.
❗Notice
All models in this series only measure vibration severity, without spectrum / envelope demodulation, and cannot perform spectrum-based fault diagnosis.
Uniform frequency response: 10Hz–1kHz, not suitable for low-frequency large units.
VB-82xx series: portable vibration meter, NOT vibration analyzer.
Vergleichstabelle: LUTRON VB-8200 / VB-8220 / VB-8206SD
Positionierung: Portable inspection vibration meters of LUTRON VB-82xx series. No spectrum analysis, vibration severity measurement only. Uniform frequency response: 10Hz~1kHz. Supplied with VB-83 separated piezoelectric probe + magnetic base, BNC interface.
| Artikel | VB‑8200 | VB‑8220 | VB‑8206SD |
| Measurable Parameters | Vibration Velocity VEL, Acceleration ACC | Vibration Velocity VEL, Acceleration ACC, Peak-to-Peak Displacement (p-p) | Vibration Velocity VEL, Acceleration ACC, Peak-to-Peak Displacement (p-p) |
| RMS / Peak | RMS / Peak | RMS / Peak | |
| Datenspeicherung | No local storage | No local storage | SD card local storage, auto logging with adjustable sampling interval; offline acquisition supported |
| Only temporary MAX/MIN record and HOLD | Only temporary MAX/MIN record and HOLD | ||
| Kommunikationsschnittstelle | RS232 serial port (real-time upload to PC) | RS232 serial port (real-time upload to PC) | RS232 serial port + SD card dual mode; data export to PC |
| Anzeige | Standard LCD | Standard LCD | Hintergrundbeleuchtetes LCD, available for low-light environment |
| Kernvorteile | Basismodell, kostengünstig; leicht, spare parts easy to source | Adds displacement measurement based on VB‑8200; displacement reading available for maintenance | SD card for long-term auto logging, continuous monitoring without permanent PC connection |
| Typical Purpose | Routine factory predictive maintenance inspection, periodic check, overhaul acceptance (vibration severity evaluation by mm/s per ISO2372) | Motor / pump overhaul, cases requiring both velocity and displacement, simple vibration severity assessment | Continuous monitoring during equipment commissioning, fault trend tracking, raw data archiving |
| Nachteile | No displacement, no SD card; only on-site reading or real-time online acquisition | No SD card; only on-site reading or online acquisition | Höherer Preis; SD card needs regular backup |
| Empfohlene Anwendungsszenarien | Most factory conventional predictive maintenance inspection | Equipment overhaul, coupling alignment inspection, when reports require displacement data | Continuous monitoring during equipment commissioning, fault trend tracking, raw data retention |
| Selection Conclusion | Preferred option if only vibration severity (mm/s) ist nötig | Select when peak-to-peak displacement is required in reports | Select when long-term auto logging and offline data storage are required |
Ergänzende Hinweise
- All three models cannot perform spectrum analysis or envelope demodulation. They only read vibration magnitude and cannot distinguish unbalance, misalignment or bearing faults. Professional vibration analyzer is required for further diagnosis once vibration exceeds limit.
- VB-83 probe is interchangeable among the three main units, magnetic base compatible.
- Stromversorgung: All use 9V PP3 battery.
Schütz,Leistungsschalter,Sensor,Encoder,SPS,Konverter