Mitutoyo HH-336 Hardness Tester - Schütz,Leistungsschalter,Sensor,Encoder,SPS,Konverter

Wechat: +86-13184948252 WhatsApp: 0086-13811255435 E-Mail: kent@bestcontactor.com

Um Kontakt |

 Mitutoyo HH-336 Hardness Tester - Schütz,Leistungsschalter,Sensor,Encoder,SPS,Konverter

Stromzähler/

Mitutoyo HH-336 Hardness Tester

Model No.: 811-336-10 (HH-336-01 is the equivalent variant) Serie: HARDMATIC HH-300 (811 Serie) Typ: Compact Digital Shore A Durometer Basic Specifications Scale: Shore A Form: Kompakte Größe, digital display Indenter: 35° truncated cone, tip diameter φ0.79 mm, shank φ1.25 mm Indenter projection: 2.5 mm Applicable materials: General rubber, elastomers, soft PVC, soft plastics, leather, neoprene ...

  • Produktdetails

Model No.: 811-336-10 (HH-336-01 is the equivalent variant)

Serie: HARDMATIC HH-300 (811 Serie)

Typ: Compact Digital Shore A Durometer

Grundlegende Spezifikationen

Skala: Shore A

Bilden: Kompakte Größe, Digitalanzeige

Indenter: 35° truncated cone, tip diameter φ0.79 mm, shank φ1.25 mm

Indenter projection: 2.5 mm

Applicable materials: General rubber, elastomers, soft PVC, soft plastics, leather, neoprene and other soft materials

Stromversorgung: SR44 silver oxide button cell

Gesamtabmessungen: 59×40×147 mm

Gewicht: 290 G

Funktionen: Peak hold, data lock, Digimatic output for MeasurLink data collection

Optionales Zubehör

Test stand for calibration: CTS-103 (Bestell-Nr. 811-013), exclusively for HH-335 / HH-336

Shore A calibration test blocks, data cable

Leitfaden zur Serienauswahl

HH-330: Compact digital Shore E (Schaum, ultra-soft rubber)

HH-332: Long-handle digital Shore A (grooves, deep cavities)

HH-334: Long-handle digital Shore D (hard rubber & rigid plastics)

HH-336: Compact digital Shore A (dieses Modell)

HH-338: Compact digital Shore D

Zusätzliche Hinweise

Compliance-Standards: ISO 7619, ASTM D2240

Compact model: for flat open specimens; long-handle model: for narrow recessed areas

Alternative: Domestic digital Shore A durometers can be used for comparison, but indent geometry and metrology systems differ. Original Mitutoyo is preferred for export quality control.

Mitutoyo HH-336 (811-336-10 Digital Shore A Durometer)

Standard: ASTM D2240 / ISO 7619; with peak hold & Digimatic data output; Batterie: SR44

  1. Full Operating Procedures (Handheld & CTS-103 Stand Mount)

Pre-Test Preparation

  1. Reinigung: Wipe the pressure foot, indenter and specimen surface free of oil, dust and debris. Never strike the indenter.
  2. Specimen requirements: Rubber/elastomer thickness ≥6 mm; measuring points ≥12 mm from edges; spacing between points ≥6 mm; place specimen on a rigid flat plate without soft backing.
  3. Einschalten: Short press the power key. With no load on the indenter, the reading shall be 0.0 HA; perform zero reset if offset.
  4. Modusauswahl: Use Peak Hold to lock the maximum reading automatically; real-time reading mode is also available.

Stand mounting (CTS-103): Fix the durometer vertically and apply load by dead weight to eliminate manual pressure variation, recommended for outgoing inspection and metrology comparison.

Handheld Measurement Steps

  1. Keep the durometer perpendicular to the specimen surface; no tilting.
  2. Press down smoothly until the entire pressure foot fully contacts the specimen (do not only push the indenter); complete the action within 1 zweite.
  3. Peak mode: The display locks the peak hardness value; record the reading and release to let the indenter rebound to zero.
  4. Take at least 5 readings per sample, discard outliers and calculate the average; increase point spacing to 15 mm for soft foams.
  5. Power off after use; remove the battery for long-term storage.

CTS-103 Stand Operation (QC Preferred)

  1. Clamp the HH-336 firmly and adjust vertical alignment; clean the worktable.
  2. Place the specimen at the table center.
  3. Turn the crank to lift the worktable slowly until the pressure foot fully seats on the specimen, then record the reading.
  4. Lower the table and repeat on new test points.

Important prohibition: Do not hold continuous pressure on the sample; soft rubber will creep and cause low readings.

  1. Kalibrierung & Verification Methods (Daily Check, Block Verification, Annual Metrology)

⚠️ The internal spring and indenter geometry are not user-adjustable. If out of tolerance, send the unit to Mitutoyo or an accredited lab for repair. The following are only verification steps, not hardware adjustment.

  1. Daily Zero Check (Every Power-On)
  2. Suspend the unit with no force on the indenter; power on and press ZERO to set the display to 0.0.
  3. Press firmly against a flat optical glass plate until the indenter retracts fully. The reading shall be approximately 100 HA (tolerance ±1 HA).

If not reaching 100 HA: Clean the indenter gap with alcohol, repeat pressing 3–5 times. If still abnormal, stop use and submit for inspection.

  1. Verification with Standard Hardness Blocks (Täglich / Batch Change)
  2. Use certified Shore A calibration blocks covering low, medium and high ranges (z.B., 30A, 60A, 80A).
  3. Zero first, then press vertically onto the block with full foot contact, record the peak reading and compare with the block nominal value.
  4. Pass criterion: Error ≤ ±1 HA. If exceeded, discontinue use and send for calibration.

Recommended cycle:

Werkstatt: Daily verification

External calibration: Once per year (CNAS certificate)

After dropping or indenter impact: Immediately stop use and re-calibrate.

  1. Full Calibration Items by Metrology Lab

Zero point & full-scale (100 HA on glass) indication error

Indenter projection length: 2.5 mm nominal

Indenter geometry: 35° cone, tip diameter φ0.79 mm

Spring test force verification

III. Fehlerbehebung

FehlerGrundursacheAbhilfe
Keine Anzeige / black screenDead SR44 battery, umgekehrte Polarität, oxidierte KontakteReplace SR44 cell; clean spring contacts; Polarität prüfen
Kann nicht auf Null gesetzt werden / zero driftSticking indenter, rubber/dust trapped in gap, Frühlingsmüdigkeit, faulty zero keyClean indenter gap with alcohol cotton swabs; do not apply oil; re-test after cleaning; persistent drift → send for repair
Unstable readings, large variation between repeated tests1. Tilting instrument; 2. Thin/soft specimen with bottom compression; 3. Incomplete foot contact; 4. Material creep; 5. Worn indenterMeasure vertically; thicken sample or use rigid backing plate; ensure full foot contact; shorten dwell time; inspect indenter tip for chipping
Low readings, cannot reach high valuesChipped/worn indenter tip, weakened spring, stuck indenterVisually inspect indenter; sauber; if no improvement, replace indenter/spring assembly at service center
Reading far below 100 HA on glassForeign matter jamming indenter, deformed pressure foot, internal mechanical damageClean thoroughly; never strike the unit; if unresolved, send to metrology service (no user spring adjustment)
Hold function fails / keys unresponsiveEindringen von Feuchtigkeit, aged switches, temporary program glitchAusschalten, remove battery and rest 5 minutes before reinstallation; keep unit dry; do not rinse with water
No Digimatic outputMismatched cable, dirty port, incorrect MeasurLink settingsUse genuine Mitutoyo Digimatic cable; clean connector pins; verify software communication parameters
Err error code on screenOverrange or mechanical jammingRelease the indenter, zero and restart; repeated errors indicate internal failure – send for service
  1. Wartung & Vorsichtsmaßnahmen
  2. Never drop or strike the indenter; the hard metal tip will be damaged permanently after impact.
  3. Only clean with anhydrous ethanol and soft cotton swabs. Do not use oil or WD40, which will trap debris and cause sticking.
  4. Avoid prolonged sustained compression to prevent spring fatigue.
  5. Lagerung: Keep dry at ambient temperature, vor direkter Sonneneinstrahlung schützen, ozone and high heat. Remove the battery during long-term storage to prevent leakage.
  6. Scale distinction: HH-336 = Shore A. Use HH-338 (Shore D) for rigid plastics; Shore A readings are invalid on hard materials.

Full Model List of Mitutoyo Hardness Testers (Nach Kategorie)

Notiz: HH-300 (811 Serie) = HARDMATIC Shore Durometers; HH-411/HH-V400 = Rebound Leeb testers; HR = Rockwell; HV = Vickers; HM = Micro-Vickers; Suffix -01 denotes equivalent variants. Stands are listed separately.

  1. HH-300 Series Shore Durometers (811 Series for Rubber / Schaum / Plastik, including HH-336)
ModellBestell-Nr.AnzeigeAnwendung
HH-329811-329-10Analog dialSchaum, ultra-soft rubber, felt, sponge
HH-330811-330-10DigitalSchaum, soft sponge, deep recesses
HH-331811-331-10Analog dialGeneral rubber, soft PVC
HH-332811-332-10DigitalGummi, soft plastic, narrow grooves
HH-333811-333-10Analog dialHard rubber, rigid plastic, bakelite
HH-334811-334-10DigitalEpoxy, hard plastic, hard rubber
HH-335811-335-10Analog dialElastomers, general rubber
HH-336811-336-10DigitalGeneral rubber & soft plastic (aktuelles Modell)
HH-337811-337-10Analog dialHard rubber, rigid plastic
HH-338811-338-10DigitalHard plastic, rigid rubber

Matching Test Stands (HH-300 Series)

CTS-102: For HH-331 / HH-333 / HH-337

CTS-103: For HH-335 / HH-336 / HH-338

  1. Portable Rebound Leeb Hardness Testers (Metal Field Inspection)

HH-411: Portable Leeb Hardness Tester

HH-V400: Upgraded rebound Leeb model

  1. HR Series Rockwell / Superficial Rockwell / Brinell Bench Hardness Testers (Metal Testing)

Classic HR-100/200/300/400 (810/963 Serie)

HR-110MR: Dial Rockwell

HR-210MR: Digital Rockwell

HR-320MS: Digital Superficial Rockwell

HR-430MR: Combined Rockwell + Superficial Rockwell

HR-430MS: Superficial Rockwell

HR-500 Series (Inner Wall / Deep Hole Special Model)

HR-530: Standard inner-wall Rockwell

HR-530L: Extended long-stroke version

HR-523 / HR-523L

HR-600 CNC Automatic Rockwell (Rockwell + Brinell for Large Workpieces)

HR-610A

HR-620A

HR-620B (Motorized Y-axis stage, fully automatic multi-point measurement)

  1. HM Series Micro-Vickers Hardness Testers (Indentations, Beschichtungen, Thin Sheets, Heat-Treated Layers)

HM-102

HM-103 (Entry-level micro-Vickers)

HM-210A / HM-210B / HM-210C / HM-210D

HM-220A / HM-220B / HM-220C / HM-220D

A = Manual; B = Auto indent reading; C = Motorized multi-point stage; D = Top model with auto-focus

  1. HV Series Vickers Hardness Testers (High-Load Vickers for Metal & Heat-Treated Parts)

HV-100 Series:

HV-110A / HV-120A: Basic touchscreen integrated models

HV-110B / HV-120B: Auto indent measurement

HV-110C / HV-120C: Multi-point measurement system

HV-110D / HV-120D: Top configuration with auto-focus

  1. Other Special Hardness Test Systems

MZT-500 Series: Material evaluation micro-hardness systems

AVK-C0: Automatic Vickers indent measuring unit

AR/ARK/ATK/ABK/ASH Series: Older legacy models (gradually discontinued)