Carlo Gavazzi VNY4 전도성 레벨 프로브 - 접촉기,회로 차단기,감지기,인코더,PLC,변환기

위챗: +86-13184948252 왓츠앱: 0086-13811255435 이메일: kent@bestcontactor.com

에 대한 연락하다 |

 Carlo Gavazzi VNY4 전도성 레벨 프로브 - 접촉기,회로 차단기,감지기,인코더,PLC,변환기

감지기/

Carlo Gavazzi VNY4 전도성 레벨 프로브

Conductive multi-point level electrode (conductivity level sensor), used for multi-point level detection of conductive liquids. Usually paired with Carlo Gavazzi level relays (such as VN-series controllers). ✅ Model Definition VNY: Y-type conductive level probe with standard nylon housing and screw terminal wiring (no pre-wired cable). 4: 4 electrode rods, enabling level detection at 4 자질 ...

  • 제품 세부정보

Conductive multi-point level electrode (conductivity level sensor), used for multi-point level detection of conductive liquids. Usually paired with 카를로 가바치 level relays (such as VN-series controllers).

✅ Model Definition

VNY: Y-type conductive level probe with standard nylon housing and screw terminal wiring (no pre-wired cable).

4: 4 electrode rods, enabling level detection at 4 measuring points.

Related models: VNY1 / VNY2 / VNY3 / VNY4; VN = cable outlet type; VNI = isolated electrode version.

📌 Key Specifications

  1. 전극: 4 AISI316 stainless steel probes, 5 mm diameter, standard length 1 m (can be cut on-site as required).
  2. 주택: Red nylon, threaded mounting. 스레드 사양: 1½” ISO228-1 pipe thread.
  3. 배선: 나사 터미널, IP67 보호 등급.
  4. Measuring principle: Conductivity type. Only suitable for conductive liquids (물, sewage, coolant, 등.); NOT for pure deionized water or oils.
  5. Medium temperature range: 0~90℃.
  6. 애플리케이션: 높은 / low level, multi-stage level alarm, water tank / sump level control.

🛠 Typical Applications

Water supply & 배수, sewage lift pits, 냉각탑, boiler water make-up, 수처리 장비

공조, conductive liquid level monitoring in food & 음료 산업, industrial tank level switches

📋 Selection Guidelines

The VNY4 is only an electrode probe. It cannot operate standalone. It must be connected to an external conductivity level controller (예를 들어. Carlo Gavazzi VN-series level relays).

Electrode length: Standard 1 m. Can be cut on-site; take care with insulating sleeves.

Medium restrictions: Oils, pure water and non-conductive liquids cannot be measured. Use special material versions for highly corrosive media.

설치: Top-mounted vertical installation. Electrodes must not touch each other and should be kept away from tank walls.

🔁 Cross-reference & 대안

VN4: Same 4-electrode design with PVC cable outlet instead of screw terminals.

VNYI4: 4-electrode isolated electrode version, preferred for scaling or interference-prone conditions.

국내 대안: Local conductive level rod (4-폴, 316엘, 1.5″ 실, IP67).

⚠️ Common Troubleshooting

  1. False switching: Electrode scaling → grind and clean; low liquid conductivity makes conductivity technology unsuitable.
  2. 신호 없음: 느슨한 배선, short circuit from electrodes touching tank wall, controller power supply fault.
  3. Corrosion: High chloride content in medium → upgrade to titanium electrode version.

Carlo Gavazzi VNY4 (4‑pole Conductivity Level Electrode) 적용사례

VNY4 advantage: 4 measuring points, which can simultaneously implement: low-low level (펌프 공회전 방지), low level (pump start), 높은 수준 (pump stop), high-high level (overflow alarm). No mechanical sticking issues as seen with float switches. Used with Carlo Gavazzi CLD4MA / CLP4MA level relays.

사례 1: HVAC Cooling Tower Make-up Water Tank (건물 / Factory HVAC)

Working condition: Cooling tower make-up tank, industrial circulating water, temperature 0~60℃. Liquid contains ions and is conductive.

4-point assignment

  1. LL Low-low level: Interlock to shut off make-up pump to prevent dry-running.
  2. L Low level: Activate make-up solenoid valve / make-up pump.
  3. H High level: Stop water make-up.
  4. HH High-high level: Acoustic & visual overflow alarm.

Selection reason: VNY4 nylon housing resists water vapour, IP67, tank-top 1½” threaded mounting. Not affected by flow turbulence unlike float switches.

메모: For heavily scaling water, prefer VNYI4 isolated electrode version to reduce false triggering.

사례 2: Sewage Lift Pit / Underground Sump (상수도 & 폐수 처리)

Working condition: Domestic sewage, impurity-laden conductive wastewater, automatic drainage control for sewage pump rooms.

4-point assignment

  1. LL: Extremely low level, block drain pump start (드라이런 보호).
  2. 엘: Low water level, start pump 1.
  3. 시간: High water level, stop pump 1; further level rise triggers standby pump 2.
  4. HH: Ultra-high level, overflow alarm, signal to PLC / building automation system.

특징: Sewage carries solids which often jam float switches. Conductive electrodes have no moving parts and longer maintenance intervals.

Risk note: Regular electrode grinding and cleaning required for scaling sewage. Upgrade to titanium electrodes for high-chloride wastewater.

사례 3: 음식 & Beverage Process Tanks (CIP Cleaning Water, Conductive Dosing Liquids)

Working condition: CIP cleaning tanks in beverage plants, ion-containing water tanks, fermenter cleaning fluids. Conductive medium, max temperature 90℃.

4-point logic: Low-low protection, low level feed start, high level feed stop, high-high overflow alarm, interlock inlet valve.

한정: Pure deionized water cannot be used. For viscous syrup media, VNYI4 isolated electrode is recommended to avoid false level readings from liquid clinging.

사례 4: Boiler Softened Water / Raw Water Pre-treatment Tanks (수처리 산업)

Working condition: Raw water tanks and softening tanks in water plants and boiler rooms, tap water / groundwater.

제어 논리: 4-stage level for automatic water make-up plus two-stage safety alarm, interlocked with PLC for pump house monitoring.

이점: AISI316 electrodes resist chloride corrosion from tap water. Probes can be cut on-site to match tank depth.

사례 5: Small Chemical Storage Tanks (Dilute Conductive Acid / Alkali Solutions)

Working condition: Dilute acid, alkali conductive waste liquid tanks, process safety interlock.

4 measuring points: Low-low level (feed pump protection), low level start feed, high level stop feed, high-high level overflow emergency inlet valve shutoff.

메모: Material assessment required for strongly corrosive media. Standard VNY4 is only suitable for weakly corrosive dilute solutions; special electrodes for heavy corrosion.

사례 6: Vehicle Wash Circulation Water Tank (Automotive Washing Machine OEM)

Working condition: Recirculating clean water tank for car wash equipment, conductive aqueous detergent solution, high vibration environment.

4 전철기: Dry-run protection, make-up start, make-up stop, overflow alarm, integrated into the wash machine control cabinet.

For heavy vibration applications, VN4 cable outlet version can replace VNY screw terminal model to prevent loose terminals.

General Selection & 애플리케이션 노트

  1. Medium must be conductive. Oils, ultra-pure water and organic solvents cannot use VNY4. Capacitance level probes are required instead.
  2. VNY4 is only a probe. Must be paired with 4-point level controllers such as CLD4MA / CLP4MA and cannot be wired directly to PLC.
  3. Metal tanks can use the tank body as common electrode. For non-metallic tanks, 그만큼 4 electrodes include the common electrode.
  4. 설치: Vertical tank-top mounting. Electrodes must be kept away from tank walls and must not short-circuit against each other.

Failure Prediction (Common On-site Issues)

Scaling: Sludge / limescale on electrode surface causes unstable level signals → regular grinding and cleaning. Use VNYI4 isolated electrode for scaling-prone media.

Low liquid conductivity: Unstable signals. Conductivity technology is not suitable; switch to ultrasonic or capacitance level meter.

Carlo Gavazzi V-series Conductive Level Probes Full Model List

Main core series: VN / VNY / VNI / VNYI (1½” 파이프 스레드, AISI316 electrodes, standard rod length 1000 mm); auxiliary compact series: VPC, VPP, 버몬트 / VTI

  1. VNY Series (Screw terminal wiring, standard non-isolated electrodes, VNY4 parent series)

VNY1: 1 electrode

VNY2: 2 전극

VNY3: 3 전극

VNY4: 4 electrodes ✅

  1. VNYI Series (나사 터미널, isolated electrode version, preferred for scaling / material clinging conditions)

VNYI1: 1 electrode

VNYI2: 2 전극

VNYI3: 3 전극

VNYI4: 4 전극

  1. VN Series (PVC cable outlet, 2 m standard cable, standard non-isolated electrodes)

VN1: 1 electrode

VN2: 2 전극

VN3: 3 전극

VN4: 4 전극

  1. VNI Series (PVC cable outlet, 2 m standard cable, isolated electrodes)

VNI1: 1 electrode

VNI2: 2 전극

VNI3: 3 전극

VNI4: 4 전극

VN / VNI cables can be extended. 예: VNI1-10M = 10-metre cable version.

  1. VPC Compact PVC Housing Series (Small threads, short electrodes, 소형 장비용)

VPC105: 1 electrode, ½” 실, 500 mm electrode length

VPC205: 2 전극, ½” 실, 500 mm electrode length

VPC110: 1 electrode, 1″ 실, 1000 mm electrode length

VPC210: 2 전극, 1″ 실, 1000 mm electrode length

VPC310: 3 전극, 1″ 실, 1000 mm electrode length

  1. VPP PVDF Insulation Corrosion-resistant Series (Small threads)

VPP105: 1 electrode, ½” 실

VPP205: 2 전극, ½” 실

VPP110: 1 electrode, 1″ 실

VPP210: 2 전극, 1″ 실

VPP310: 3 전극, 1″ 실

  1. 버몬트 / VTI Teflon Insulated Series (Strongly corrosive applications)

VT1 / VTI1: 1 electrode

VT2 / VTI2: 2 전극

VT3 / VTI3: 3 전극

VT4 / VTI4: 4 전극

VT = standard; VTI = isolated electrode; 1 m silicone cable outlet.

Series Comparison Table

시리즈콘센트 유형Electrode Insulation주택실일반적인 응용
VNY나사식 터미널기준나일론 61½”Water tanks, 냉각탑, local cabinet wiring (VNY4)
VNYI나사식 터미널PE isolated나일론 61½”Sewage, scaling-prone or viscous conductive liquids
VN2m PVC cable기준나일론 61½”Tank top where cable routing is easier than terminal wiring
VNI2m PVC cablePE isolated나일론 61½”Scaling applications with direct cable exit
VPC나사식 터미널체육PVC½”/1″소형 기계, compact tanks
VPP나사식 터미널PVDFPP½”/1″Mildly corrosive media
VT/VTI1m silicone cableTeflonTeflon1½”Dilute acid / alkali chemical corrosion

Matching Controllers (필수적인; probe cannot work alone)

CLD4MA, CLP4MA (4-point level relays, compatible with VNY4 / VNYI4 / VN4 / VNI4); CLD2MA for 2-electrode probes; CLD3MA for 3-electrode probes.

Naming Rule Summary

  1. VNY = V+NY (NY = Nylon + 나사식 터미널, nylon housing with screw terminals)
  2. VN = V+N (nylon housing, 케이블 콘센트)
  3. Suffix I = Isolated electrode (prevents false triggering from liquid film clinging)
  4. Final digit = number of electrodes (1/2/3/4)