EL3020 belongs to the EasyLine EL3000 series. It is the 19-inch rack-mount version in АББ’s environmental analysis product line (the wall-mounted version of the same series is EL3040).
Основные характеристики
Монтаж: 19-inch rack, 3U height
Размеры: 483 мм (Вт) × 357 мм (Д) × 132 мм (ЧАС)
Класс защиты: IP20; IP40 for CEMS version
Источник питания: 100~240 VAC
Структура: A single cabinet can be fitted with a maximum of 2 analysis modules, measuring up to 5 gas components simultaneously
Дополнительные аксессуары: Built-in sample gas pump, электромагнитные клапаны, фильтр, flow sensor
Источник: Германия
Available Measuring Modules (mix and match supported)
| Модель модуля | Принцип | Measurable Gas |
| Uras26 | NDIR Non-Dispersive Infrared | СО, CO₂, CH₄, SO₂, НЕТ |
| Limas23 | UV Ultraviolet Absorption | НЕТ, NO₂, SO₂ |
| Magnos28 | Paramagnetic Oxygen | O₂ |
| Magnos27 | Thermomagnetic Oxygen | O₂ |
| ZO23 | Zirconia Trace Oxygen | Trace O₂ |
| Caldos27 | Thermal Conductivity | Н₂, Ar and other binary gas mixtures |
| Fidas24 | FID Flame Ionization Detection | Total Hydrocarbons / HC |
Интерфейсы & Коммуникация
Ethernet (встроенный веб-сервер)
Профибус ДП
Модбус
4…20 mA analog output
Digital DI/DO (Макс. 12 ОТ + 12 ДЕЛАТЬ)
Local LCD with 5-button operation, Chinese display supported; ECT configuration software for advanced parameter setup
Ключевые особенности
- Built-in calibration cell: Clean air can be used for zero calibration in most working conditions, eliminating the need for permanent standard gas cylinders; automatic calibration supported
- Built-in self-diagnostics, optional QAL3 (EN14181) certification for flue gas continuous monitoring, suitable for CEMS emission monitoring
- ABB Ability digital function: Quick QR code access for diagnostics, remote support, status alerts, cluster monitoring on cloud platform
- Модульная конструкция: Modules can be added or replaced later without replacing the whole unit
Типичные применения
Flue gas emission monitoring (CEMS) for power plants, waste incineration, metallurgy and boilers
Biogas, landfill gas and fermentation gas analysis
Blast furnace gas and reformer process gas analysis
Gas production, air separation, inert gas purity and trace oxygen monitoring
Combustion monitoring for engines and gas turbines
Запчасти & Customs Information
Код ТН ВЭД: 90279000 (same classification for gas analyzer spare parts and complete unit)
Common consumables: Sample filter element, pump diaphragm, FID flame nozzle, gas capillary, FFKM sealing O-ring, источник света
Quick Comparison with EL3040
EL3020: Rack-mounted, IP20, installed inside control cabinet
EL3040: Настенный, IP65 purge type, field local installation
Operating Instructions for ABB EL3020 EasyLine Extractive Analyzer
The front panel adopts 5-button operation: ХОРОШО (confirm / войти в меню), вверх / вниз / левый / right navigation keys. The screen supports Chinese language. The operation modes include measurement view, menu setting, calibration and diagnosis. It can also be operated remotely via ECT configuration software on PC or built-in web page.
Important precondition: EL3020 is 19-inch rack-mounted with IP20, no cabinet purge. Remove the yellow protective cap of the pressure sensor before power-on. Check gas circuit, heat tracing, фильтр, pump and flow in advance; no condensation or leakage allowed.
- Power-on Operation (First Startup / Restart after Long Shutdown)
- Pre-startup inspection
Gas circuit: Sample gas, zero gas and span gas pipelines connected properly, switching valves functional; filter cartridge intact; flow rate set to rated value (generally 30–60 L/h)
Electrical circuit: 100–240 VAC power supply, 4–20 мА, DI/DO and communication wiring correct
Pressure sensor: Remove yellow plastic cap and connect T-type pressure compensation pipeline as required
- Pipeline pre-purge (mandatory for new commissioning or after maintenance)
Purge the whole sample gas loop with nitrogen or clean zero gas to remove residual impurities and corrosive gas and avoid contamination of optical modules. Purge time depends on pipeline length, normally ≥20 min.
- Power on and warm-up
Switch on power; the analyzer starts automatically. Warm-up duration varies by module
Uras26 infrared: приблизительно. 30 мин
Limas23 UV, Magnos paramagnetic oxygen: 30–60 min
Fidas24 FID: ≥60 min, requires hydrogen and combustion air
ZO23 zirconia trace oxygen: longer thermostatic stabilization time
- After warm-up completion, the main measurement interface appears, showing component concentration, единица, поток, status and alarm information. Sample gas can be fed in for continuous measurement.
Кончик: Versions with built-in calibration cell (Uras26) can use clean air for daily zero calibration without continuous zero gas cylinder supply.
- Panel Buttons & Menu Access
Normal measurement screen: Real-time concentration, статус, flow and temperature
Press【OK】to enter main menu; select items with up/down keys, confirm with OK, modify parameters with left/right keys, exit with return key
Main menu hierarchy:
- Measurement View
- Калибровка (most frequently used)
- Диагностика / Тревога
- Конфигурация: диапазон, единица, порог тревоги, аналоговый выход, коммуникация, время, language
- Бревно: журнал тревог, calibration record, QAL3 report (необязательный)
- Info: module version, serial number
- Daily Operation Monitoring (Normal Continuous Measurement)
- Feed stable sample gas after pre-treatment, filtration and moisture removal before entering the analyzer
- Screen observation:
Stable readings for each component, no flashing fault codes
Flow within set range; low flow triggers alarm to protect gas cell
Status indicator: зеленый = нормальный; yellow = warning; красный = неисправность
- Remote access:
Enter analyzer IP address in browser to open built-in web server for real-time data, кривые, alarms and calibration records. Use ECT EasyLine Configuration Tool for advanced parameter setting via RS232 service port or Ethernet. TCT software is used for in-depth diagnosis of optical modules.
- Выходной сигнал: 4–20 мА, Модбус / Profibus transmit concentration, status and fault signals to PLC/DCS synchronously.
- Калибровка (Ключ: Ноль / Span)
Принцип: Zero first, then span; start calibration only after readings stabilize with gas flowing. Follow QAL3 cycle requirements for CEMS environmental applications.
4.1 Manual Zero Calibration
- Menu path: OK → Calibration → Zero Cal
- Select target channel (CO/SO₂/NO/O₂ etc.)
- Switch gas circuit to zero gas (high-purity N₂ or clean zero air free of target components). Keep gas flowing for 3–5 min until reading stable.
- Select【Start Zero Calibration】
- Wait for automatic completion; screen shows Zero OK and reading returns near zero
- Switch back to sample gas
4.2 Manual Span Calibration
- Menu path: OK → Calibration → Span Cal
- Select target channel and input nominal concentration on gas certificate
- Switch to standard span gas, keep gas flowing for 5–10 min until reading stable
- Start span calibration; the instrument automatically corrects slope
- Completion prompt Span OK, check deviation between reading and certified span gas concentration
- Switch back to sample gas for measurement
4.3 Automatic Calibration (Models with built-in calibration cell, recommended for long-term operation)
- Меню: Calibration → Auto Cal for setting
- Set calibration cycle, последовательность, automatic valve switching and alarm logging
- After enabling auto-calibration, the analyzer drives internal solenoid valves at scheduled time and completes zero/span via built-in calibration cell, no external span gas cylinder required (standard for Uras26)
- FID, Limas, ZO23 and Caldos modules do NOT support built-in calibration cell; external standard gas is mandatory.
Calibration Differences by Module
Uras26 (NDIR): Supports auto zero via built-in cell; span can use internal cell or external standard gas
Limas23 (УФ): Generally calibrated with external standard gas
Magnos28 paramagnetic oxygen: Single-point calibration with clean air
Fidas24 FID THC: Zero with zero air, span with methane standard gas; stable hydrogen pressure required
ZO23 trace oxygen: Calibrated with reference air; pay attention to sealing against leakage
- Тревога & Сброс неисправности
- Alarm checking: Menu Diagnosis for active alarms and historical alarm log with text description
Common alarms: low flow, temperature abnormality, light source failure, pressure fault, calibration failure, FID flameout, over-range
- Reset after troubleshooting:
General warning: auto-recover after root cause eliminated
Latched fault: select【Reset Fault】in menu; some severe faults require power cycle
- Важный: After FID flameout, check hydrogen and combustion air, purge before re-ignition; avoid repeated ignition directly.
- Конфигурация параметров
- Range modification: Set upper/lower limit for each component; recalibration is recommended after range change
- 4–20 mA analog output: Set mA mapping range, fault current (3.5 мА / 22 мА)
- DI/DO: Custom digital input (external calibration trigger, блокировка) and digital output (вина, calibration in progress, high concentration alarm)
- Коммуникация: Modbus TCP/RTU, Profibus address and baud rate; Modbus port is disabled by factory default and needs enabling via ECT
- Язык, time and password management (advanced parameters require engineer password)
- Shutdown Operation
- Short shutdown (within several hours): Cut sample gas supply, keep purge gas and power on with constant temperature, no power-off to prevent condensate contamination of gas cell
- Long shutdown (>1 day)
Purge gas circuit with zero gas for 5–10 min
For FID version: Cut hydrogen first, extinguish flame, continue purging FID with combustion air
Close sample gas and standard gas cylinders
Finally cut off analyzer power supply
Seal pipelines for dust prevention
- Daily Operation Notes
- Sample gas must be pre-treated: dewatering, dust removal and tar removal; condensate is strictly forbidden to enter the analyzer
- Record each calibration: время, span gas concentration and result, archive for CEMS QAL3 compliance
- Периодическая проверка: filter element, pump diaphragm, flow and pressure; check FID nozzle and hydrogen pressure reducing valve regularly
- Avoid long-time impact of over-range or high-concentration corrosive gas which may damage light source and gas cell
- Troubleshooting for failed calibration: gas leakage, unstable flow, expired standard gas, water vapor interference or light source aging
- Brief ECT Software Operation on PC
- Связь: RS232 service port or Ethernet
- Read device parameters and back up configuration
- Modify advanced calibration parameters, curves and compensation coefficients (давление & температурная компенсация)
- Firmware upgrade and deep diagnosis (TCT software for optical module light source and gas cell diagnosis)
ABB EasyLine EL3000 Series Complete Model & Информация для заказа
Важное примечание:
EL3000 does not have fixed single part numbers. It is a configurable system consisting of chassis + measuring module + ввод/вывод + gas accessories. The complete system code is a long string `24042-xxxxxxxxxx` generated according to configuration and cannot be fixed statically. Below are base chassis models, measuring module models, descriptions, sample system codes and spare part ordering numbers.
There are only two base chassis models in this series: EL3020 (19” rack, IP20) and EL3040 (wall-mounted IP65 purge type). EL3010 / EL3030 / EL3050 do not exist. EL3060 is an independent Magnos206 unit and NOT part of the EL3020 / EL3040 chassis family.
- Base Chassis Models (2 Типы)
| Chassis Model | Монтаж | Защита | Описание |
| EL3020 | 19-inch rack, 3ты | IP20 (IP40 for CEMS version) | Installed in control cabinet; optional built-in sample pump, solenoid valves and filter; Макс. 2 модули, до 5 компоненты |
| EL3040 | Настенный | IP65 (purge capable) | Field local installation; purge housing for dust and water protection; also supports max. 2 модули |
- All Compatible Measuring Modules for EL3000 Series
| Модель модуля | Принцип | Measurable Gas | Функции |
| Uras26 | NDIR Non-Dispersive Infrared | СО, CO₂, SO₂, НЕТ, CH₄, N₂O | With built-in calibration cell, mainstream for CEMS, до 4 infrared components |
| Limas23 | UV Ultraviolet Absorption | SO₂, НЕТ, NO₂ | Flue gas UV module; no built-in calibration cell, external standard gas required |
| Magnos206 | Paramagnetic Oxygen | O₂ | New paramagnetic oxygen, replacement for legacy Magnos28; range down to 0–2%Vol, supports 98–100% high oxygen |
| Magnos27 | Thermomagnetic Oxygen | O₂ | Thermomagnetic principle, for conventional oxygen, starting from 0–5%Vol |
| ZO23 | Zirconia Trace Oxygen | Trace O₂ | ppb-level trace oxygen, for inert gas and high-purity gas monitoring |
| Caldos27 | Thermal Conductivity | Н₂, Ar, Он, binary mixtures | Hydrogen and inert gas purity analysis |
| Fidas24 | FID Flame Ionization | THC / VOC | Organic carbon measurement; requires hydrogen and combustion air |
| Electrochemical Oxygen Module | Electrochemical | O₂ | Low-cost economical oxygen sensor, for non-environmental CEMS |
Combination rule: One EL3020 / EL3040 chassis can mount max. 2 modules freely combined (например. Uras26+Magnos206; Uras26+Limas23; Caldos27+ZO23)
- Sample Complete System Codes (Magnos206 Example for Reference Format)
Формат: `24042-XXXXXXXXXX` (system number)
EL3020 + Magnos206, IP20, 2 К, no flow/pressure sensor: 24042-111011000000
Matching housing: 24342-111011000001; electronic board: 24442-1100110000N1; measuring module: 24641-1110110500H1
EL3040 + Magnos206, IP65, 2 К, no flow/pressure sensor: 24042-121031000000
Matching housing: 24342-121031000001; electronic board: 24442-1200310000N1; measuring module: 24641-1210310500Л1
The prefix for other combinations (Uras26, Limas23, Fidas24 etc.) is identical. Middle digits change with module type, ввод/вывод, источник питания, сертификация (QAL3) and built-in pump/valve. Formal ordering code must be issued by ABB sales based on final configuration.
- Common Spare Part Order Numbers (Compatible with EL3020 / EL3040)
| Номер детали | Описание | Applicable Module |
| 769305 | Modulated light source assembly (dual radiator CaF₂) | Uras26 |
| 769342 | Combined gas capillary assembly (without catalyst) | EL3020 complete unit |
| 9AAC176461 | Fidas24 piping kit | Fidas24 |
| 9AAC176378 | Uras26 modulation unit | Uras26 |
Other consumables: pump diaphragm, FFKM O-ring, filter cartridge, FID nozzle, flow sensor and pressure sensor use separate short codes extracted according to module revision.
- Выбор & Ordering Rules for ABB Inquiry
- Шасси: EL3020 / EL3040
- Модуль 1 модель + диапазон + component; Модуль 2 модель + диапазон + component (None if only one module)
- Источник питания: 100–240 VAC
- ввод/вывод: quantity of 4–20 mA AO, DI/DO
- Коммуникация: Modbus TCP/RTU, Профибус-ДП
- Built-in gas circuit: sample pump, solenoid valve bank, filter or not
- Сертификация: QAL3 (EN14181) for CEMS or not
- Аксессуары: flow sensor, pressure compensation sensor, calibration valve bank
- Distinguish Easily Confused Models
EL3000 = series collective name
EL3020 = 19-inch rack chassis
EL3040 = wall-mounted IP65 chassis
EL3060 = standalone Magnos206 paramagnetic oxygen analyzer (not EL3020/EL3040 chassis module)
Uras26 / Limas23 / Magnos206 / Fidas24 etc.: measuring modules, not complete analyzer models
- Дополнительные примечания
- No public full fixed part number master list, as it is configurable. Any change in range, ввод/вывод, certification or gas accessories will alter the trailing code.
- I can provide an EL3000 inquiry template (Excel field list) for you to send to ABB distributors for formal code and quotation.
- Separate spare parts lists for Uras26, Limas23 and Fidas24 are also available on request.
Контактор,автоматический выключатель,датчик,Кодер,ПЛК,Конвертер

















NH42-63-318x560.png)



