Model No.: 811-336-10 (HH-336-01 is the equivalent variant)
Série: HARDMATIC HH-300 (811 Série)
Taper: Compact Digital Shore A Durometer
Spécifications de base
Scale: Shore A
Form: Taille compacte, affichage numérique
Indenter: 35° truncated cone, tip diameter φ0.79 mm, shank φ1.25 mm
Indenter projection: 2.5 mm
Applicable materials: General rubber, elastomers, soft PVC, soft plastics, leather, neoprene and other soft materials
Alimentation: SR44 silver oxide button cell
Dimensions hors tout: 59×40×147 mm
Poids: 290 g
Fonctions: Peak hold, data lock, Digimatic output for MeasurLink data collection
Accessoires en option
Test stand for calibration: CTS-103 (Numéro de commande. 811-013), exclusively for HH-335 / HH-336
Shore A calibration test blocks, data cable
Series Selection Guide
HH-330: Compact digital Shore E (mousse, ultra-soft rubber)
HH-332: Long-handle digital Shore A (grooves, deep cavities)
HH-334: Long-handle digital Shore D (hard rubber & rigid plastics)
HH-336: Compact digital Shore A (ce modèle)
HH-338: Compact digital Shore D
Notes complémentaires
Normes de conformité: OIN 7619, ASTM D2240
Compact model: for flat open specimens; long-handle model: for narrow recessed areas
Alternative: Domestic digital Shore A durometers can be used for comparison, but indent geometry and metrology systems differ. Original Mitutoyo is preferred for export quality control.
Mitutoyo HH-336 (811-336-10 Digital Shore A Durometer)
Standard: ASTM D2240 / OIN 7619; with peak hold & Digimatic data output; batterie: SR44
- Full Operating Procedures (Handheld & CTS-103 Stand Mount)
Pre-Test Preparation
- Nettoyage: Wipe the pressure foot, indenter and specimen surface free of oil, dust and debris. Never strike the indenter.
- Specimen requirements: Rubber/elastomer thickness ≥6 mm; measuring points ≥12 mm from edges; spacing between points ≥6 mm; place specimen on a rigid flat plate without soft backing.
- Mise sous tension: Short press the power key. With no load on the indenter, the reading shall be 0.0 HA; perform zero reset if offset.
- Sélection des modes: Use Peak Hold to lock the maximum reading automatically; real-time reading mode is also available.
Stand mounting (CTS-103): Fix the durometer vertically and apply load by dead weight to eliminate manual pressure variation, recommended for outgoing inspection and metrology comparison.
Handheld Measurement Steps
- Keep the durometer perpendicular to the specimen surface; no tilting.
- Press down smoothly until the entire pressure foot fully contacts the specimen (do not only push the indenter); complete the action within 1 deuxième.
- Peak mode: The display locks the peak hardness value; record the reading and release to let the indenter rebound to zero.
- Take at least 5 readings per sample, discard outliers and calculate the average; increase point spacing to 15 mm for soft foams.
- Power off after use; remove the battery for long-term storage.
CTS-103 Stand Operation (QC Preferred)
- Clamp the HH-336 firmly and adjust vertical alignment; clean the worktable.
- Place the specimen at the table center.
- Turn the crank to lift the worktable slowly until the pressure foot fully seats on the specimen, then record the reading.
- Lower the table and repeat on new test points.
Important prohibition: Do not hold continuous pressure on the sample; soft rubber will creep and cause low readings.
- Étalonnage & Méthodes de vérification (Daily Check, Block Verification, Annual Metrology)
⚠️ The internal spring and indenter geometry are not user-adjustable. If out of tolerance, send the unit to Mitutoyo or an accredited lab for repair. The following are only verification steps, not hardware adjustment.
- Daily Zero Check (Every Power-On)
- Suspend the unit with no force on the indenter; power on and press ZERO to set the display to 0.0.
- Press firmly against a flat optical glass plate until the indenter retracts fully. The reading shall be approximately 100 HA (tolerance ±1 HA).
If not reaching 100 HA: Clean the indenter gap with alcohol, repeat pressing 3–5 times. If still abnormal, stop use and submit for inspection.
- Verification with Standard Hardness Blocks (Tous les jours / Batch Change)
- Use certified Shore A calibration blocks covering low, medium and high ranges (par ex., 30UN, 60UN, 80UN).
- Zero first, then press vertically onto the block with full foot contact, record the peak reading and compare with the block nominal value.
- Pass criterion: Error ≤ ±1 HA. If exceeded, discontinue use and send for calibration.
Recommended cycle:
Atelier: Daily verification
External calibration: Once per year (CNAS certificate)
After dropping or indenter impact: Immediately stop use and re-calibrate.
- Full Calibration Items by Metrology Lab
Point zéro & full-scale (100 HA on glass) indication error
Indenter projection length: 2.5 mm nominal
Indenter geometry: 35° cone, tip diameter φ0.79 mm
Spring test force verification
III. Dépannage
| Faute | Cause première | Remède |
| Pas d'affichage / black screen | Dead SR44 battery, polarité inversée, contacts oxydés | Replace SR44 cell; clean spring contacts; check polarity |
| Cannot zero / zero drift | Sticking indenter, rubber/dust trapped in gap, fatigue printanière, faulty zero key | Clean indenter gap with alcohol cotton swabs; do not apply oil; re-test after cleaning; persistent drift → send for repair |
| Unstable readings, large variation between repeated tests | 1. Tilting instrument; 2. Thin/soft specimen with bottom compression; 3. Incomplete foot contact; 4. Material creep; 5. Worn indenter | Measure vertically; thicken sample or use rigid backing plate; ensure full foot contact; shorten dwell time; inspect indenter tip for chipping |
| Low readings, cannot reach high values | Chipped/worn indenter tip, weakened spring, stuck indenter | Visually inspect indenter; faire le ménage; if no improvement, replace indenter/spring assembly at service center |
| Reading far below 100 HA on glass | Foreign matter jamming indenter, deformed pressure foot, internal mechanical damage | Clean thoroughly; never strike the unit; if unresolved, send to metrology service (no user spring adjustment) |
| Hold function fails / keys unresponsive | Pénétration d'humidité, aged switches, temporary program glitch | Éteindre, remove battery and rest 5 minutes before reinstallation; keep unit dry; do not rinse with water |
| No Digimatic output | Mismatched cable, dirty port, incorrect MeasurLink settings | Use genuine Mitutoyo Digimatic cable; nettoyer les broches du connecteur; verify software communication parameters |
| Err error code on screen | Overrange or mechanical jamming | Release the indenter, zero and restart; repeated errors indicate internal failure – send for service |
- Entretien & Précautions
- Never drop or strike the indenter; the hard metal tip will be damaged permanently after impact.
- Only clean with anhydrous ethanol and soft cotton swabs. Do not use oil or WD40, which will trap debris and cause sticking.
- Avoid prolonged sustained compression to prevent spring fatigue.
- Stockage: Keep dry at ambient temperature, à l'abri de la lumière directe du soleil, ozone and high heat. Remove the battery during long-term storage to prevent leakage.
- Scale distinction: HH-336 = Shore A. Use HH-338 (Shore D) for rigid plastics; Shore A readings are invalid on hard materials.
Full Model List of Mitutoyo Hardness Testers (Par catégorie)
Note: HH-300 (811 Série) = HARDMATIC Shore Durometers; HH-411/HH-V400 = Rebound Leeb testers; HR = Rockwell; HV = Vickers; HM = Micro-Vickers; suffixe -01 denotes equivalent variants. Stands are listed separately.
- HH-300 Series Shore Durometers (811 Series for Rubber / Foam / Plastique, including HH-336)
| Modèle | Numéro de commande. | Afficher | Application |
| HH-329 | 811-329-10 | Analog dial | Foam, ultra-soft rubber, felt, sponge |
| HH-330 | 811-330-10 | Numérique | Foam, soft sponge, deep recesses |
| HH-331 | 811-331-10 | Analog dial | General rubber, soft PVC |
| HH-332 | 811-332-10 | Numérique | Caoutchouc, soft plastic, narrow grooves |
| HH-333 | 811-333-10 | Analog dial | Hard rubber, rigid plastic, bakelite |
| HH-334 | 811-334-10 | Numérique | Epoxy, hard plastic, hard rubber |
| HH-335 | 811-335-10 | Analog dial | Elastomers, general rubber |
| HH-336 | 811-336-10 | Numérique | General rubber & soft plastic (modèle actuel) |
| HH-337 | 811-337-10 | Analog dial | Hard rubber, rigid plastic |
| HH-338 | 811-338-10 | Numérique | Hard plastic, rigid rubber |
Matching Test Stands (HH-300 Series)
CTS-102: For HH-331 / HH-333 / HH-337
CTS-103: For HH-335 / HH-336 / HH-338
- Portable Rebound Leeb Hardness Testers (Metal Field Inspection)
HH-411: Portable Leeb Hardness Tester
HH-V400: Upgraded rebound Leeb model
- HR Series Rockwell / Superficial Rockwell / Brinell Bench Hardness Testers (Metal Testing)
Classic HR-100/200/300/400 (810/963 Série)
HR-110MR: Dial Rockwell
HR-210MR: Digital Rockwell
HR-320MS: Digital Superficial Rockwell
HR-430MR: Combined Rockwell + Superficial Rockwell
HR-430MS: Superficial Rockwell
HR-500 Series (Inner Wall / Deep Hole Special Model)
HR-530: Standard inner-wall Rockwell
HR-530L: Extended long-stroke version
HR-523 / HR-523L
HR-600 CNC Automatic Rockwell (Rockwell + Brinell for Large Workpieces)
HR-610A
HR-620A
HR-620B (Motorized Y-axis stage, fully automatic multi-point measurement)
- HM Series Micro-Vickers Hardness Testers (Indentations, Revêtements, Thin Sheets, Heat-Treated Layers)
HM-102
HM-103 (Entry-level micro-Vickers)
HM-210A / HM-210B / HM-210C / HM-210D
HM-220A / HM-220B / HM-220C / HM-220D
A = Manual; B = Auto indent reading; C = Motorized multi-point stage; D = Top model with auto-focus
- HV Series Vickers Hardness Testers (High-Load Vickers for Metal & Heat-Treated Parts)
HV-100 Series:
HV-110A / HV-120A: Basic touchscreen integrated models
HV-110B / HV-120B: Auto indent measurement
HV-110C / HV-120C: Multi-point measurement system
HV-110D / HV-120D: Top configuration with auto-focus
- Other Special Hardness Test Systems
MZT-500 Series: Material evaluation micro-hardness systems
AVK-C0: Automatic Vickers indent measuring unit
AR/ARK/ATK/ABK/ASH Series: Older legacy models (gradually discontinued)
Contacteur,disjoncteur,capteur,Encodeur,API,Convertisseur




















NH42-63-318x560.png)